• Journal of Infrared and Millimeter Waves
  • Vol. 37, Issue 3, 302 (2018)
HUANG Fei1、2、3、*, LI Fan-Ming1、3, WANG Shi-Yong1、3, CHEN Chao-Shuai1、2、3, and SHI Yong-Biao1、2、3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less
    DOI: 10.11972/j.issn.1001-9014.2018.03.009 Cite this Article
    HUANG Fei, LI Fan-Ming, WANG Shi-Yong, CHEN Chao-Shuai, SHI Yong-Biao. Calibration method for NIR polarization detector[J]. Journal of Infrared and Millimeter Waves, 2018, 37(3): 302 Copy Citation Text show less

    Abstract

    In this paper, a new NIR polarization detector is presented. It is fabricated out of a wire grid polarizer attaching to the surface of a FPA. A brand new method solving linear polarization is proposed at the same time. The transmittance and the angle of the wire grid are corrected. The effect on the degree of linear polarization (DoLP) measurement is also eliminated in this report. Theoretically, the infrared polarized image quality is independent of the transmittance and angle of wire grid. The experiment uses a blackbody source as the calibration light. The results show that, while ignoring the crosstalk between the polarized pixels, the measured value is about 14%~24% of the true value. After adding the transmittance correction, the data reaches the level of 80%~120% while it becomes 96% to 104% by applying the angle correction.
    HUANG Fei, LI Fan-Ming, WANG Shi-Yong, CHEN Chao-Shuai, SHI Yong-Biao. Calibration method for NIR polarization detector[J]. Journal of Infrared and Millimeter Waves, 2018, 37(3): 302
    Download Citation