• Chinese Journal of Lasers
  • Vol. 42, Issue 5, 516003 (2015)
Xu Danyang1、*, Tong Jianping1, Gao Jianxun1, and Wang Fei2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/cjl201542.0516003 Cite this Article Set citation alerts
    Xu Danyang, Tong Jianping, Gao Jianxun, Wang Fei. Fiber Spectrometer Optical Simulation Optimization and Calibration[J]. Chinese Journal of Lasers, 2015, 42(5): 516003 Copy Citation Text show less

    Abstract

    The optical system is the core part of a spectrometer, and it determines the basic properties and the volume of the entire instrument. Advantages and disadvantages of several optical systems are analyzed, and the asymmetric Czerny-Turner system is adopted. The resulting optical path optimized with Zemax has an overall resolution of 1.8 nm, and a measurement range of 200~900 nm. Relative error of wavelength is controlled to be below 0.05 nm, with cubic spline interporlation coefficients obtained from Matlab and HG- 1 mercury- argon used as the calibration light source. Analyses on data for comparision with the USB4000 fiber spectrometer show that both of the design and implementation are feasible.
    Xu Danyang, Tong Jianping, Gao Jianxun, Wang Fei. Fiber Spectrometer Optical Simulation Optimization and Calibration[J]. Chinese Journal of Lasers, 2015, 42(5): 516003
    Download Citation