• Journal of Atmospheric and Environmental Optics
  • Vol. 18, Issue 1, 82 (2023)
HU Rongchun1,2, LU Fan1,*, and ZHOU Peisong1
Author Affiliations
  • 1Key Laboratory of Environmental Optics and Technology, Anhui Institute of Optics and Fine Mechanics, HFIPS,Chinese Academy of Sciences, Hefei 230031, China
  • 2University of Science and Technology of China, Hefei 230026, China
  • show less
    DOI: 10.3969/j.issn.1673-6141.2023.01.008 Cite this Article
    Rongchun HU, Fan LU, Peisong ZHOU. Design of hydrogen flame ionization detection system[J]. Journal of Atmospheric and Environmental Optics, 2023, 18(1): 82 Copy Citation Text show less
    Principle diagram of hydrogen flame ionization detection system
    Fig. 1. Principle diagram of hydrogen flame ionization detection system
    Circuit block diagram of the system
    Fig. 2. Circuit block diagram of the system
    Current detection diagram. (a) Current noise; (b) linear curve and detection range
    Fig. 3. Current detection diagram. (a) Current noise; (b) linear curve and detection range
    Experiments of methane standard gases. (a) Chromatogram; (b) linear curve
    Fig. 4. Experiments of methane standard gases. (a) Chromatogram; (b) linear curve
    被测电流/pA反馈电阻/Ω理论输出电压/mV测量输出电压/mV相对误差/%
    0.11 G0.10.0982.01
    11 G1.00.9821.81
    1001 G100.099.2110.80
    2001 G200.0198.3230.85
    5001 G500.0499.4210.12
    10001 G1000.0998.6230.14
    100001 M10.010.1311.01
    1000001 M100.0100.5120.50
    2000001 M200.0200.6210.30
    5000001 M500.0500.3210.06
    10000001 M1000.0999.9870.01
    Table 1. Current detection error
    Sample volume/μLPeak area/(pA·s)RSD/%
    12345
    5072.1373.6772.3671.4970.851.47
    200250.22255.72252.16255.56253.280.92
    500521.03512.13507.32515.53525.221.37
    Table 2. Repeatability of the system