• Optical Instruments
  • Vol. 41, Issue 5, 10 (2019)
Lei ZHANG, Xinzhi SHAN, and Xiumin GAO*
Author Affiliations
  • School of Optical-Electrical and Computer Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China
  • show less
    DOI: 10.3969/j.issn.1005-5630.2019.05.002 Cite this Article
    Lei ZHANG, Xinzhi SHAN, Xiumin GAO. Non-invasive blood pressure detection system based on pulse wave[J]. Optical Instruments, 2019, 41(5): 10 Copy Citation Text show less
    Vascular model
    Fig. 1. Vascular model
    Systematic hardware
    Fig. 2. Systematic hardware
    Blood pressure arithmetic circuit and calibration circuit
    Fig. 3. Blood pressure arithmetic circuit and calibration circuit
    The flow chart of the system’s software
    Fig. 4. The flow chart of the system’s software
    The flow chart of data acquisition program
    Fig. 5. The flow chart of data acquisition program
    Comparison of blood pressure test data
    Fig. 6. Comparison of blood pressure test data
    被测者编号测量值/mmHg标准值/mmHg绝对误差/mmHg相对误差/%
    A1 126/74125/711/30.7/4.1
    A2 121/79121/780/10/1.3
    B1 127/73128/70−1/30.8/4.1
    B2 126/76124/692/71.6/9.2
    C1 116/68120/72−4/−43.4/5.9
    C2 118/65119/70−1/−50.8/7.7
    D1 126/65128/63−2/21.6/3.1
    D2 125/75121/724/33.2/4
    E1 123/77124/79−1/−20.8/2.6
    E2 127/73122/745/−13.9/1.4
    F1 131/82126/795/33.8/3.7
    F2 129/80128/781/−20.8/2.5
    G1 124/76128/73−4/30.8/3.9
    G2 127/73129/68−2/51.6/6.8
    H1 119/76124/77−5/−14.2/1.3
    H2 121/79118/753/42.5/5.1
    I1 112/62118/65−6/−35.4/4.8
    I2 115/65119/60−4/53.5/7.7
    J1 135/83129/806/34.4/3.6
    J2 131/79132/81−1/−20.8/2.5
    Table 1. Comparisons between blood pressure test values and standard values