• Chinese Journal of Lasers
  • Vol. 50, Issue 17, 1714017 (2023)
Zhuorui Zheng1、4, Hui Zhong1、*, Yongxiao Nie1、2, Ting Lin1、3, Yifei Fang1, Liwei Song1、4, and Ye Tian1、4
Author Affiliations
  • 1State Key Laboratory of High Field Laser Physics,Shanghai Institute of Optics and Fine Mechanics,Chinese Academy of Sciences,Shanghai 201800,China
  • 2School of Optical and Electronic Information, Huazhong University of Science and Technology, Wuhan 430074, Hubei, China
  • 3College of Optical-Electrical and Computer Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China
  • 4Center of Materials Science and Optoelectronics Engineering,University of Chinese Academy of Sciences,Beijing 100049,China
  • show less
    DOI: 10.3788/CJL230881 Cite this Article Set citation alerts
    Zhuorui Zheng, Hui Zhong, Yongxiao Nie, Ting Lin, Yifei Fang, Liwei Song, Ye Tian. Research of Bovine Serum Albumin Sensor Based on Terahertz Metamaterials[J]. Chinese Journal of Lasers, 2023, 50(17): 1714017 Copy Citation Text show less
    Schematic of experimental set up with micrographs of metamaterials without determinand and with dry determinand shown in inset
    Fig. 1. Schematic of experimental set up with micrographs of metamaterials without determinand and with dry determinand shown in inset
    Transmission spectra before and after placing metamaterials in optical path
    Fig. 2. Transmission spectra before and after placing metamaterials in optical path
    Transmittance curves and corresponding Gaussian fitting curves under different volume mass values of determinands. (a) 2 mg/mL; (b) 3 mg/mL; (c) 4 mg/mL; (d) 8 mg/mL
    Fig. 3. Transmittance curves and corresponding Gaussian fitting curves under different volume mass values of determinands. (a) 2 mg/mL; (b) 3 mg/mL; (c) 4 mg/mL; (d) 8 mg/mL
    Sensor resonance frequency offset versus volume mass of determinand
    Fig. 4. Sensor resonance frequency offset versus volume mass of determinand
    Transmittance curves and corresponding Gaussian fitting curves under different volume mass values of determinands. (a) 1.0 mg/mL; (b) 0.3 mg/mL; (c) 0.2 mg/mL
    Fig. 5. Transmittance curves and corresponding Gaussian fitting curves under different volume mass values of determinands. (a) 1.0 mg/mL; (b) 0.3 mg/mL; (c) 0.2 mg/mL

    Volume mass /

    (mg·mL-1

    Resonance peak without

    determinand /GHz

    Resonance peak with

    determinand /GHz

    f /GHz
    2804.8798794.444910.4349
    3805.2425792.937912.3046
    4800.5992786.522514.0767
    8802.9098782.558520.3513
    Table 1. Resonance frequencies and frequency offsets under different volume mass values of determinands

    Volume mass /

    (mg·mL-1

    Resonance peak

    without determinand /GHz

    Resonance peak

    with determinand /GHz

    f /GHz
    1.0800.8577798.85372.0040
    0.3800.1836799.31070.8729
    0.2801.6216801.58310.0385
    Table 2. Resonance frequencies and frequency offsets under different volume mass values of determinands
    Zhuorui Zheng, Hui Zhong, Yongxiao Nie, Ting Lin, Yifei Fang, Liwei Song, Ye Tian. Research of Bovine Serum Albumin Sensor Based on Terahertz Metamaterials[J]. Chinese Journal of Lasers, 2023, 50(17): 1714017
    Download Citation