• Spectroscopy and Spectral Analysis
  • Vol. 32, Issue 1, 41 (2012)
ZHAO Hui*, WANG Gui-shi, CAI Ting-dong, and GAO Xiao-ming
Author Affiliations
  • [in Chinese]
  • show less
    DOI: 10.3964/j.issn.1000-0593(2012)01-0041-05 Cite this Article
    ZHAO Hui, WANG Gui-shi, CAI Ting-dong, GAO Xiao-ming. Off-Axis Cavity Enhanced Absorption Spectroscopy Detection Techniques for the Measurement of Carbon Dioxide[J]. Spectroscopy and Spectral Analysis, 2012, 32(1): 41 Copy Citation Text show less

    Abstract

    The spectrum of carbon monoxide was obtained around 1.573 μm using a tunable distributed feedback semiconductor laser with a high-finesse cavity at room temperature via off-axis cavity enhanced absorption (CEA) spectroscopic technique. The absorption line of carbon monoxide at 6 357.311 6 cm-1 was chosen for trace detection. Meanwhile, in order to get more accurate measurements, absorption path length of the cavity calibration methods was studied, and a simple and practical calibration method was given. The results show that, the equivalent absorption path length of high-precision optical resonator was ~1 195.73 m. At last, we got the concentration of carbon monoxide in the real atmosphere to be ~388.346 ppm (S/N≈22), and the detection limit of carbon monoxide was 17.65 ppm. By combination of wavelength modulation technology and OA-CEAS technology, a minimum detectable concentration of 0.36 ppm (S/N≈1 064) was achieved eventually.
    ZHAO Hui, WANG Gui-shi, CAI Ting-dong, GAO Xiao-ming. Off-Axis Cavity Enhanced Absorption Spectroscopy Detection Techniques for the Measurement of Carbon Dioxide[J]. Spectroscopy and Spectral Analysis, 2012, 32(1): 41
    Download Citation