• Acta Photonica Sinica
  • Vol. 47, Issue 8, 847003 (2018)
FEI Teng1、*, PAN Cong-yuan2, ZENG Qiang1, DU Xue-wei1, and WANG Qiu-ping1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/gzxb20184708.0847003 Cite this Article
    FEI Teng, PAN Cong-yuan, ZENG Qiang, DU Xue-wei, WANG Qiu-ping. Absolute Calibration of Laser-induced Breakdown Spectroscopy Detection System by Using a Tungsten Halogen Lamp[J]. Acta Photonica Sinica, 2018, 47(8): 847003 Copy Citation Text show less

    Abstract

    An absolute calibration setup for laser-induced breakdown spectroscopy detection system is built based on the radiometry theory. A tungsten halogen lamp equipped with an ultraviolet glass filter and a fused silica diffuser is used as standard light source for the calibration. Absolute calibration of a laser-induced breakdown spectroscopy detection system with a Czerny-Turner spectrometer in the ultraviolet range is performed. The absolute spectral response of the system in the range of 310~385 nm is obtained with an uncertainty less than 7% by 2 standard deviation estimate. The absolute spectral calibration provides a way to evaluate the hardware of the laser-induced breakdown spectroscopy detection system.
    FEI Teng, PAN Cong-yuan, ZENG Qiang, DU Xue-wei, WANG Qiu-ping. Absolute Calibration of Laser-induced Breakdown Spectroscopy Detection System by Using a Tungsten Halogen Lamp[J]. Acta Photonica Sinica, 2018, 47(8): 847003