• Chinese Optics Letters
  • Vol. 20, Issue 8, 081202 (2022)
Shaowen Zhang1、2、3, Jixi Lu2、4、*, Ying Zhou2、**, Fei Lu1, Kaifeng Yin1, Di Zhan1, Yueyang Zhai2、4, and Mao Ye2、4
Author Affiliations
  • 1School of Instrumentation and Optoelectronic Engineering, Beihang University, Beijing 100191, China
  • 2Hangzhou Innovation Institute of Beihang University, Hangzhou 310051, China
  • 3Beihang Hangzhou Innovation Institute Yuhang, Hangzhou 310023, China
  • 4Research Institute for Frontier Science, Beihang University, Beijing 100191, China
  • show less
    DOI: 10.3788/COL202220.081202 Cite this Article Set citation alerts
    Shaowen Zhang, Jixi Lu, Ying Zhou, Fei Lu, Kaifeng Yin, Di Zhan, Yueyang Zhai, Mao Ye. Zero field optically pumped magnetometer with independent dual-mode operation[J]. Chinese Optics Letters, 2022, 20(8): 081202 Copy Citation Text show less

    Abstract

    We propose a dual-mode optically pumped magnetometer (OPM) that can flexibly switch between single-beam modulation mode and double-beam DC mode. Based on a 4 mm×4 mm×4 mm miniaturized vapor cell, the double-beam DC mode achieves a sensitivity of 7 fT/Hz1/2 with probe noise below 4 fT/Hz1/2 and working bandwidth over 65 Hz. This mode is designed to precisely measure the noise floor of a mu-metal magnetic shield. The single-beam modulation mode (sensitivity 20 fT/Hz1/2) exhibits bandwidth characteristics suitable for biomagnetic measurements. Thus, our design is suitable for a miniaturized OPM with multiple functions, including magnetic-shield background noise measurement and medical imaging.
    Pt=1q[γeB×P+ROP(sz^P)RrelP],

    View in Article

    Rrel=RSD+1TD+1T2SE,

    View in Article

    Px=0,

    View in Article

    Py=γeROP(γeBx)2+(ROP+Rrel)2Bx,

    View in Article

    Pz=ROP(ROP+Rrel)(γeBx)2+(ROP+Rrel)2.

    View in Article

    Pz(t)=kROPJ0(γeB1ω)J1(γeB1ω)(γeBx)2+(ROP+Rrel)2sin(ωt)γeBx.

    View in Article

    Psingle or doubleROP(ROP+Rrel)2γeBx.

    View in Article

    k1=η1IprobelnrecfD1γe,

    View in Article

    k2=vprobevD1(vprobevD1)2+(ΓD1/2)2eOD(vprobe),

    View in Article

    k3=ROP(ROP+Rrel)2,

    View in Article

    vprobevD1=±ΓD12OD(vD1)+OD2(vD1)+12.

    View in Article

    Ipump=hvpumprecfD1(vpumpvD1)2+(ΓD1/2)2ΓD1/2(RSD+1TD).

    View in Article

    k4=η2Ipumpγe,

    View in Article

    k5=e(1pz)OD(vpump),

    View in Article

    k6=ROPJ0(γeB1ω)J1(γeB1ω)(ROP+Rrel)2,

    View in Article

    Shaowen Zhang, Jixi Lu, Ying Zhou, Fei Lu, Kaifeng Yin, Di Zhan, Yueyang Zhai, Mao Ye. Zero field optically pumped magnetometer with independent dual-mode operation[J]. Chinese Optics Letters, 2022, 20(8): 081202
    Download Citation