• NUCLEAR TECHNIQUES
  • Vol. 45, Issue 10, 100502 (2022)
Peng NIE*, Jin LU, Yaping GUO, Ren REN, and Lijun ZHANG
Author Affiliations
  • China Institute of Atomic Energy, Beijing 102413, China
  • show less
    DOI: 10.11889/j.0253-3219.2022.hjs.45.100502 Cite this Article
    Peng NIE, Jin LU, Yaping GUO, Ren REN, Lijun ZHANG. Study on approach for measurement of radioactive waste in steel box based on ISOCS[J]. NUCLEAR TECHNIQUES, 2022, 45(10): 100502 Copy Citation Text show less
    In situ objects counting system
    Fig. 1. In situ objects counting system
    Ring collimator
    Fig. 2. Ring collimator
    Diagram showing steel box measurement using ISOCS
    Fig. 3. Diagram showing steel box measurement using ISOCS
    The decay of the γ ray through the absorber
    Fig. 4. The decay of the γ ray through the absorber
    Simulation curve interface for the efficiency calibration without a radioactive source
    Fig. 5. Simulation curve interface for the efficiency calibration without a radioactive source
    Geometry information for steel box
    Fig. 6. Geometry information for steel box
    Using ISOCS for efficiency calibration
    Fig. 7. Using ISOCS for efficiency calibration
    Schematic diagram of simulation volume source efficiency calibration with point source
    Fig. 8. Schematic diagram of simulation volume source efficiency calibration with point source
    Diagram showing the distribution of the standard radioactive source
    Fig. 9. Diagram showing the distribution of the standard radioactive source
    序号Serial number源位置Source position

    137Cs放射性活度

    Radioactivity of 137Cs / Bq

    60Co放射性活度

    Radioactivity of 60Co / Bq

    1位置1 Position 12.22×1072.06×104
    2位置2 Position 22.04×1072.59×104
    3位置3 Position 31.93×1081.94×105
    4位置4 Position 42.00×1081.81×105
    5位置5 Position 51.05×1083.73×105
    6位置6 Position 61.55×1083.88×105
    平均测量值Average values1.16×1081.97×105
    活度标准值Standard value of activity9.01×1071.51×105
    相对偏差Relative deviation28%30%
    Table 1. Comparison between measured (average over 6 positions in a steel box) and standard values for a 137Cs and a 60Co source
    Peng NIE, Jin LU, Yaping GUO, Ren REN, Lijun ZHANG. Study on approach for measurement of radioactive waste in steel box based on ISOCS[J]. NUCLEAR TECHNIQUES, 2022, 45(10): 100502
    Download Citation