• Chinese Journal of Lasers
  • Vol. 20, Issue 3, 171 (1993)
[in Chinese]1, [in Chinese]1, [in Chinese]1, [in Chinese]1, [in Chinese]1, and [in Chinese]2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: Cite this Article Set citation alerts
    [in Chinese], [in Chinese], [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Measurements of refractive index of lithium niobate integrated optical substrate[J]. Chinese Journal of Lasers, 1993, 20(3): 171 Copy Citation Text show less

    Abstract

    The measurments of refractive index of lithium niobate integrated optical substrates were performed by means of prism coupling method on a precision goniometer with He-Ne laser. In the experiment, we used the differential technique and a liquid crystal optical rotator. The result shows that the acuracy is of 1×10-4 and the reproducible deviations are less than 5×10-5.
    [in Chinese], [in Chinese], [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Measurements of refractive index of lithium niobate integrated optical substrate[J]. Chinese Journal of Lasers, 1993, 20(3): 171
    Download Citation