    Lin LIU, Yang XIANG, Lixin LIU, Weidong AN, Lei HUANG. Simple medical rigid endoscopic relative percent distortion detection system[J]. Journal of Applied Optics, 2023, 44(4): 756 Copy Citation Text show less
    Schematic diagram of 30° laparoscope
    Fig. 1. Schematic diagram of 30° laparoscope
    Relative percent distortion curve of endoscope
    Fig. 2. Relative percent distortion curve of endoscope
    Pattern of calibration target plate
    Fig. 3. Pattern of calibration target plate
    Schematic diagram of calibration
    Fig. 4. Schematic diagram of calibration
    Distortion of target plate
    Fig. 5. Distortion of target plate
    Schematic diagram of overall system design
    Fig. 6. Schematic diagram of overall system design
    Flow chart of system work
    Fig. 7. Flow chart of system work
    Physical image of overall testing device
    Fig. 8. Physical image of overall testing device
    Diagram of software interface
    Fig. 9. Diagram of software interface
    Coincidence images of calibration target plate
    Fig. 10. Coincidence images of calibration target plate
    Calibrated relative percent distortion measurement image
    Fig. 11. Calibrated relative percent distortion measurement image
    Relative percent distortion detection images without vertical calibration
    Fig. 12. Relative percent distortion detection images without vertical calibration
    次数方式方块位置平均值方差
    1校准24.124.824.323.224.10.6
    2校准24.424.324.523.924.30.2
    3校准23.824.424.323.624.00.3
    4未校准24.727.128.823.626.12.0
    5未校准27.224.524.129.826.42.3
    6未校准24.128.129.224.826.62.2
    Table 1. Measurement results of relative percent distortion of endoscope%
    Lin LIU, Yang XIANG, Lixin LIU, Weidong AN, Lei HUANG. Simple medical rigid endoscopic relative percent distortion detection system[J]. Journal of Applied Optics, 2023, 44(4): 756