• Spectroscopy and Spectral Analysis
  • Vol. 29, Issue 4, 1127 (2009)
ZHANG Nan1、*, SHENG Qing-hai2, SHI Zhen-hua1, ZHANG Zhi-guo2, DUAN Xiang-lin1, and CHANG Yan-zhong1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: Cite this Article
    ZHANG Nan, SHENG Qing-hai, SHI Zhen-hua, ZHANG Zhi-guo, DUAN Xiang-lin, CHANG Yan-zhong. Application of Solid Sampling Graphite Furnace Atomic Absorption Spectrophotometry to Mensuration of Brain Iron Content in Rats[J]. Spectroscopy and Spectral Analysis, 2009, 29(4): 1127 Copy Citation Text show less

    Abstract

    In the present study, the authors performed the solid sampling and detected the iron levels in cortex, hippocampus and striatum of rat brain by GFAAS. The authors’ results showed that there are no remarkable difference between the data obtained by solid sampling graphite furnace atomic absorption and liquid sampling graphite furnace atomic absorption. Compared to liquid sampling graphite furnace atomic absorption, the sample pre-treatment stage was obviously simplified, the cost was reduced significantly, and the time was shortened significantly in the solid sampling GFAAS. This study will be beneficial to the mensuration of iron content in micro-tissue of animal by solid sampling GFASS.
    ZHANG Nan, SHENG Qing-hai, SHI Zhen-hua, ZHANG Zhi-guo, DUAN Xiang-lin, CHANG Yan-zhong. Application of Solid Sampling Graphite Furnace Atomic Absorption Spectrophotometry to Mensuration of Brain Iron Content in Rats[J]. Spectroscopy and Spectral Analysis, 2009, 29(4): 1127
    Download Citation