• Chinese Journal of Lasers
  • Vol. 9, Issue 3, 186 (1982)
[in Chinese]
Author Affiliations
  • [in Chinese]
  • show less
    DOI: Cite this Article Set citation alerts
    [in Chinese]. [J]. Chinese Journal of Lasers, 1982, 9(3): 186 Copy Citation Text show less

    Abstract

    In our interference length measuring system the minimum display value attainable is 0.1 μm, with a measuring length of less than 4.8 metres. The necessary correction for laser wavelength in our system is affected by interpolation of value (1) in association with the changes of correction factor C3 (on the dial) according to testing conditions of temperature and pressure. The accuracy attainable is better than 1×10-7.