• Spectroscopy and Spectral Analysis
  • Vol. 37, Issue 10, 3133 (2017)
ZHU Gao-feng1、2、*, YANG Chun-hua1, ZHU Hong-qiu1, and GUI Wei-hua1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3964/j.issn.1000-0593(2017)10-3133-05 Cite this Article
    ZHU Gao-feng, YANG Chun-hua, ZHU Hong-qiu, GUI Wei-hua. Oxygen Concentration Detection and Calibration Method Improvement in Pharmaceutical Vial Based on Wavelength Modulation Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2017, 37(10): 3133 Copy Citation Text show less

    Abstract

    Pharmaceutical vial’s oxygen concentration detection based on wavelength modulation spectroscopy (WMS) was conducted on a single, short optical path in the open. The oxygen absorption lines at 760.885 nm (13 142.58 cm-1 ) were chosen- and the system parameters were optimized. Then, data processing method and steps were explained, such as real-time background deduction and real-time spectrum correction. Seven samples of the different oxygen concentrations were tested, and the corresponding second harmonic signals were obtained. Finally, the linear regression equations were established between the concentration and the peak height(C—H), the peak area at half maximum(C—A). The experimental results indicated that the fitting coefficients of two methods were 0.996 6 and 0.997 8, respectively, and the latter was increased by 0.12 % than the former. Prediction accuracy of the model was evaluated by fully interactive authentication method and the results indicated that the root mean square errors of prediction (RMSEP) were 0.003 1 and 0.002 0, respectively, and the latter was reduced by 37.69% than the former. The 2f signal of sample ( 4%) were severally measured for 20 times by the two methods in different time and the results indicated that standard deviations were 0.002 2 and 0.001 6, respectively, and the latter was reduced by 27.3% than the former. Meanwhile, the resolutions were 0.198% and 0.097%, respectively, and the latter was increased by 51% than the former. The system is feasible to measure the pharmaceutical vial’s oxygen concentration, and the accuracy and stability of the system can be improved by the C—A method.
    ZHU Gao-feng, YANG Chun-hua, ZHU Hong-qiu, GUI Wei-hua. Oxygen Concentration Detection and Calibration Method Improvement in Pharmaceutical Vial Based on Wavelength Modulation Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2017, 37(10): 3133
    Download Citation