• High Power Laser and Particle Beams
  • Vol. 36, Issue 12, 126001 (2024)
Li An, Jun Xiao*, Xinhua Wang, Lei Xie..., Li Jiang, Jiecheng Yang, Haiping Guo and Zijie Han|Show fewer author(s)
Author Affiliations
  • Institute of Nuclear Physics and Chemistry, CAEP, Mianyang 621900, China
  • show less
    DOI: 10.11884/HPLPB202436.240320 Cite this Article
    Li An, Jun Xiao, Xinhua Wang, Lei Xie, Li Jiang, Jiecheng Yang, Haiping Guo, Zijie Han. Calibration of 6Li atomic number of lithium glass detector[J]. High Power Laser and Particle Beams, 2024, 36(12): 126001 Copy Citation Text show less

    Abstract

    In experimental validations of fusion blanket neutronics performance, tritium production rate is one of the most crucial measurement parameters. The number of 6Li atoms in the detector, as a normalization factor for calculating the tritium production rate, is a key factor determining the accuracy of measurement results and must be accurately calibrated. This paper specifically introduces the principles of calibrating the number of 6Li, the experimental setup and procedure, and the methods for quantifying uncertainties. For the first time, the number of 6Li atoms in a small lithium glass detector was calibrated using a germanium monocrystal monochromator to obtain 32.36 meV neutrons at the M5 horizontal channel of the China Mianyang Research Reactor (CMRR), with an uncertainty of 2.62%.
    Li An, Jun Xiao, Xinhua Wang, Lei Xie, Li Jiang, Jiecheng Yang, Haiping Guo, Zijie Han. Calibration of 6Li atomic number of lithium glass detector[J]. High Power Laser and Particle Beams, 2024, 36(12): 126001
    Download Citation