    Qi Zhang, Limin Zhang, Jiayu Wang, Jingyi An, Jing Wang, chen Chen, Feng Gao. Spatially Resolved Reflectance in Sub-Diffusive Regime Measurement System for Mucosal Lesion Recognition[J]. Chinese Journal of Lasers, 2021, 48(20): 2007001 Copy Citation Text show less
    Schematic of experimental system
    Fig. 1. Schematic of experimental system
    System stability measurement results
    Fig. 2. System stability measurement results
    Experimental multi-frequency crosstalk ratio
    Fig. 3. Experimental multi-frequency crosstalk ratio
    Intensity amplitude measured at different ambient light intensities. (a) Without digital lock-in; (b) digital lock-in
    Fig. 4. Intensity amplitude measured at different ambient light intensities. (a) Without digital lock-in; (b) digital lock-in
    Linearity assessments of the system at source-detector separation of 220 μm
    Fig. 5. Linearity assessments of the system at source-detector separation of 220 μm
    Optical parameter inversion results. (a) μa; (b) μ's; (c)γ
    Fig. 6. Optical parameter inversion results. (a) μa; (b) μ's; (c)γ
    Wavelength /nmSource-dectetor separation /μm
    2204406608801100
    5200.99920.99820.99770.99870.9984
    6500.99910.99940.99940.99920.9992
    7850.99880.99890.99910.99730.9985
    8300.99850.99930.99850.99880.9983
    Table 1. Correlation coefficient of linear fitting at four wavelengths
    No.μa /cm-1μ's /cm-1γNo.μa /cm-1μ's /cm-1γ
    P10.00192.2507P70.5132.0110
    P20.001132.0110P80.1171.9992
    P30.001171.9992P90.5171.9992
    P40.001292.0110P101.5171.9992
    P50.192.2507P110.1292.0110
    P62.092.2507P122.0292.0110
    Table 2. Optical parameters of turbid phantoms
    ParameterRoot-mean-square errorRelative root-mean-square error /%
    μa0.102326.0074
    μ's0.54422.9835
    γ0.03931.9139
    Table 3. Root-mean-square error and relative root-mean-square error of optical parameters