• Acta Photonica Sinica
  • Vol. 51, Issue 11, 1130001 (2022)
Lei LI, Liping TANG, Qiuyang MA, Zijiang GAO, Yang GAO, and Yingying QIAO*
Author Affiliations
  • School of Physics and Microelectronics,Zhengzhou University,Zhengzhou 450001,China
  • show less
    DOI: 10.3788/gzxb20225111.1130001 Cite this Article
    Lei LI, Liping TANG, Qiuyang MA, Zijiang GAO, Yang GAO, Yingying QIAO. CO Detection Based on Photoacoustic Spectroscopy with CEEMDAN[J]. Acta Photonica Sinica, 2022, 51(11): 1130001 Copy Citation Text show less
    Simulation of noise reduction performance
    Fig. 1. Simulation of noise reduction performance
    CO absorption coefficient diagram
    Fig. 2. CO absorption coefficient diagram
    Differential photoacoustic cell operating in a first-order longitudinal resonance mode
    Fig. 3. Differential photoacoustic cell operating in a first-order longitudinal resonance mode
    Optimized modulation amplitude
    Fig. 4. Optimized modulation amplitude
    The PAS system and differential photoacoustic cell
    Fig. 5. The PAS system and differential photoacoustic cell
    The IMF components
    Fig. 6. The IMF components
    Noise reduction performance
    Fig. 7. Noise reduction performance
    The relation between gas concentrations and second harmonic signal amplitudes(inset shows the noise measured at the pure N2 background)
    Fig. 8. The relation between gas concentrations and second harmonic signal amplitudes(inset shows the noise measured at the pure N2 background)
    PerformanceContaminated signalWaveletCEEMDAN
    SNR10.0910.9818.17
    SSIM0.230.530.66
    Table 1. Performance comparison of wavelet algorithm and the noise reduction and reconstruction algorithm based on CEEMDAN
    ComponentD1D2D3D4D5D6D7
    Ci/Cr0.2190.244-0.2110.1630.079-0.0780.036
    ComponentD8D9D10D11D12D13Dr
    Ci/Cr0.0850.6310.6590.5440.646-0.4810.078
    Table 2. The correlation coefficients of the components and original signal
    Signal amplitude / mVStandard deviation / mVSNRMDL /(×10-6
    Original signal676.35.275128.211.7
    With Wavelet652.63.134213.67.0
    With CEEMDAN669.31.114585.82.6
    Table 3. Performance parameters of PAS system
    Lei LI, Liping TANG, Qiuyang MA, Zijiang GAO, Yang GAO, Yingying QIAO. CO Detection Based on Photoacoustic Spectroscopy with CEEMDAN[J]. Acta Photonica Sinica, 2022, 51(11): 1130001
    Download Citation