• Chinese Optics Letters
  • Vol. 17, Issue 4, 041201 (2019)
Minghui Chen1、2, Wenyu Jia1, Jintao He1, Martial Geiser3, and Gang Zheng1、*
Author Affiliations
  • 1Institute of Biomedical Optics and Optometry, Shanghai Institute for Minimally Invasive Therapy, University of Shanghai for Science and Technology, Shanghai 200093, China
  • 2Beckman Laser Institute and Center for Biomedical Engineering, University of California, Irvine, CA 92612, USA
  • 3Systems Engineering Institute, University of Applied Sciences Western Switzerland, Sion 1950, Switzerland
  • show less
    DOI: 10.3788/COL201917.041201 Cite this Article Set citation alerts
    Minghui Chen, Wenyu Jia, Jintao He, Martial Geiser, Gang Zheng. A miniaturized system for measurement of the refractive index of sub-microliter liquid[J]. Chinese Optics Letters, 2019, 17(4): 041201 Copy Citation Text show less

    Abstract

    This study introduced the research and development of a portable and miniaturized system for the measurement of the refractive index of sub-microliter liquid based on a microfluidic chip. A technical method of double-beam interference, was proposed for use in the measurement. Based on this, by using a laser diode as a light source, changes in the refractive index were calculated by utilizing a complementary metal–oxide–semiconductor to detect the movement of interference fringes of the liquid. Firstly, this study simulated the effects of influencing factors on the interference infringes of two Gaussian beams, such as their spot sizes, distance between two beam spots, and detection range. Secondly, this research introduced the system design and construction of the double-beam interference method and analyzed the results of refractive index tests on sub-microliter aqueous glucose solutions with different concentrations. The measurement accuracy reached 10 4 refractive index units. This system has a compact structure and is rendered portable by using batteries for its power supply. The entire system is designed to be a double Z-shaped structure with a length of about 15 cm, a width of 5 cm, and a height of about 10 cm. It can be used to measure changes in the refractive index of sub-microliter to nanoliter liquids based on the use of a microfluidic chip.
    U1=q0q0+zexp(ikz)exp{ik2[(x+d(n)2)2+y2]q0+z},(1)

    View in Article

    U2=q0q0+zexp(ikz)exp{ik2[(xd(n)2)2+y2]q0+z},(2)

    View in Article

    I=(U1+U2)·conj(U1+U2).(3)

    View in Article

    I=2q0q0+z·coskxd(n)2(q0+z).(4)

    View in Article

    ΔL=0.44λ2Δλ,(5)

    View in Article

    Minghui Chen, Wenyu Jia, Jintao He, Martial Geiser, Gang Zheng. A miniaturized system for measurement of the refractive index of sub-microliter liquid[J]. Chinese Optics Letters, 2019, 17(4): 041201
    Download Citation