• Journal of Atmospheric and Environmental Optics
  • Vol. 10, Issue 4, 333 (2015)
Jianjun LI1、*, Fangang MENG1, Maopeng XIA1, Zhengqiang LI2, Wenchao ZHAI1, Donghui LI2, Lei DING1, Xiaobing ZHENG1, and Tao QI1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3969/j.issn.1673-6141.2015.04.008 Cite this Article
    LI Jianjun, MENG Fangang, XIA Maopeng, LI Zhengqiang, ZHAI Wenchao, LI Donghui, DING Lei, ZHENG Xiaobing, QI Tao. Laboratory Calibration Method of Sunphotometer[J]. Journal of Atmospheric and Environmental Optics, 2015, 10(4): 333 Copy Citation Text show less

    Abstract

    Sunphotometer is the most important observation instrument in global AERONET site, which supply a lot of observation data for obtaining time and space distributions of global aerosol. A kind of high accuracy calibration method against cryogenic radiometer is developed in laboratory, calibration coefficient of sun channel and sky diffuser channel can be obtained in the ambient condition of laboratory. The calibration principle and installation are introduced in detail, calibration curve of absolute spectral responsivity are obtained in major spectral channel, and measurement uncertainty is analysed. The result indicated that the uncertainty of calibration coefficient of sky diffuser channel is less than 0.8% and sun channel is less than 2.06%.
    LI Jianjun, MENG Fangang, XIA Maopeng, LI Zhengqiang, ZHAI Wenchao, LI Donghui, DING Lei, ZHENG Xiaobing, QI Tao. Laboratory Calibration Method of Sunphotometer[J]. Journal of Atmospheric and Environmental Optics, 2015, 10(4): 333
    Download Citation