• Chinese Journal of Lasers
  • Vol. 39, Issue 1, 104001 (2012)
Xie Wenming1、2、3、*, Zeng Zhiping1、2、3, Li Li1、2、3, Chen Shuqiang4, Li Zhifang1、2、3, and Li Hui1、2、3
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • 4[in Chinese]
  • show less
    DOI: 10.3788/cjl201239.0104001 Cite this Article Set citation alerts
    Xie Wenming, Zeng Zhiping, Li Li, Chen Shuqiang, Li Zhifang, Li Hui. Signal Compensation in Long-Focal-Zone Photoacoustic Imaging Systems[J]. Chinese Journal of Lasers, 2012, 39(1): 104001 Copy Citation Text show less

    Abstract

    Signal attenuation is highly concerned in many research fields. Imaging quality and depth in photoacoustic (PA) imaging system are often limited by the attenuation of PA signal. The PA signal attenuation is induced by many factors in PA imaging system. The factors leading to the signal attenuation and its characteristics are discussed based on tissue optics, acoustic transport and detection in a long-focal-zone PA imaging system. Considering to recover the PA signal from optical to ultrasonic attenuation, a method to compensate the detected PA signals is presented. The proposed method is evaluated by simulant sample and employed to image a thyroid tissue in vitro. The experimental results demonstrate that the method could be used to recover the PA signal and to improve the imaging depth and quality in the PA system.
    Xie Wenming, Zeng Zhiping, Li Li, Chen Shuqiang, Li Zhifang, Li Hui. Signal Compensation in Long-Focal-Zone Photoacoustic Imaging Systems[J]. Chinese Journal of Lasers, 2012, 39(1): 104001
    Download Citation