• Acta Physica Sinica
  • Vol. 68, Issue 7, 070703-1 (2019)
Hua-Wei Jin1、2、3, Ren-Zhi Hu1、2、*, Pin-Hua Xie1、2, Hao Chen1、2, Zhi-Yan Li1、2, Feng-Yang Wang1、2, Yi-Hui Wang1、2, and Chuan Lin1、2
Author Affiliations
  • 1Key Laboratory of Environmental Optics and Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China
  • 2University of Science and Technology of China, Hefei 230026, China
  • 3School of Mechanical Engineering, Anhui University of Science and Technology, Huainan 232001, China
  • show less
    DOI: 10.7498/aps.68.20182262 Cite this Article
    Hua-Wei Jin, Ren-Zhi Hu, Pin-Hua Xie, Hao Chen, Zhi-Yan Li, Feng-Yang Wang, Yi-Hui Wang, Chuan Lin. Photo-acoustic technology applied to ppb level NO2 detection by using low power blue diode laser [J]. Acta Physica Sinica, 2019, 68(7): 070703-1 Copy Citation Text show less
    Schematic diagram of NO2-PAS systemNO2-PAS系统示意图
    Fig. 1. Schematic diagram of NO2-PAS system NO2-PAS系统示意图
    Cross sections of NO2, H2O and diode laser spectrumNO2和H2O的吸收截面以及蓝光二极管激光光谱
    Fig. 2. Cross sections of NO2, H2O and diode laser spectrum NO2和H2O的吸收截面以及蓝光二极管激光光谱
    Frequency response of the photo-acoustic cell光声池的频率响应
    Fig. 3. Frequency response of the photo-acoustic cell光声池的频率响应
    Impact analysis of surface treatment and study of performance optimization表面处理影响分析及性能优化研究
    Fig. 4. Impact analysis of surface treatment and study of performance optimization表面处理影响分析及性能优化研究
    Comparative study of background noise between one buffer and two buffer单缓冲和两级缓冲对本底噪声的对比研究
    Fig. 5. Comparative study of background noise between one buffer and two buffer单缓冲和两级缓冲对本底噪声的对比研究
    Performance evaluationof the system系统性能评估
    Fig. 6. Performance evaluationof the system系统性能评估
    Analysis diagram of Allan varianceAllan方差分析图
    Fig. 7. Analysis diagram of Allan varianceAllan方差分析图
    Air intake system of atmospheric NO2 concentrations环境大气NO2的进气系统
    Fig. 8. Air intake system of atmospheric NO2 concentrations 环境大气NO2的进气系统
    (a) Simultaneous measurement of atmospheric NO2 concentrations by PAS and CRDS systems; (b) correlation between the atmospheric NO2 concentrations measured by PAS and CRDS systems(a) PAS系统和CRDS系统测得的环境大气NO2浓度; (b)PAS系统和CRDS系统NO2测量结果对比
    Fig. 9. (a) Simultaneous measurement of atmospheric NO2 concentrations by PAS and CRDS systems; (b) correlation between the atmospheric NO2 concentrations measured by PAS and CRDS systems (a) PAS系统和CRDS系统测得的环境大气NO2浓度; (b)PAS系统和CRDS系统NO2测量结果对比
    明细样气响应/ ${\text{μV}}$本底噪声/ ${\text{μV}}$
    腔体内表面处理前94.33 ± 2.473.27 ± 0.34
    腔体内表面处理后110.27 ± 1.042.86 ± 0.33
    光学窗片处理前3.28 ± 0.34
    光学窗片处理后3.13 ± 0.14
    线性电源2.8 ± 0.16
    Table 1.

    Comparison of test results.

    测试结果比较

    Hua-Wei Jin, Ren-Zhi Hu, Pin-Hua Xie, Hao Chen, Zhi-Yan Li, Feng-Yang Wang, Yi-Hui Wang, Chuan Lin. Photo-acoustic technology applied to ppb level NO2 detection by using low power blue diode laser [J]. Acta Physica Sinica, 2019, 68(7): 070703-1
    Download Citation