  • Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, MD 20993, USA2 Office of Device Evaluation, Center for Devices and Radiological Health, U.S. Food and Drug Administration, 9200 Corporate Boulevard, Rockville, MD 20850, USA
  • show less
    DOI: 10.3788/COL20080612.0876 Cite this Article Set citation alerts
    Ilko K., Robert W., Do-Hyun Kim, Robert H., Don Calogero. Innovative confocal laser method for exact dioptric power measurement of intraocular lens implants[J]. Chinese Optics Letters, 2008, 6(12): 876 Copy Citation Text show less

    Abstract

    We present a novel confocal laser method (CLM) for precise testing of the dioptric power of both positive and negative intraocular lens (IOL) implants. The CLM principle is based on a simple fiber-optic confocal laser design including a single-mode fiber coupler that serves simultaneously as a point light source used for formation of a collimated Gaussian laser beam, and as a highly sensitive confocal point receiver. The CLM approach provides an accurate, repeatable, objective, and fast method for IOL dioptric power measurement over the range from 0 D to greater than +-30 D under both dry and in-situ simulated conditions.