• Acta Photonica Sinica
  • Vol. 45, Issue 6, 604001 (2016)
HU You-bo1、2、*, LI Jian-jun2, XIA Mao-peng2, GAO Dong-yang2, and ZHENG Xiao-bing2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/gzxb20164506.0604001 Cite this Article
    HU You-bo, LI Jian-jun, XIA Mao-peng, GAO Dong-yang, ZHENG Xiao-bing. Measurement and Correction of Linearity of Photon Counters[J]. Acta Photonica Sinica, 2016, 45(6): 604001 Copy Citation Text show less

    Abstract

    Linear measurement device was set up based on the linear relationship between pump power and photons in spontaneous parametric down conversion process. The nonlinearity errors mechanism, linear condition and correction model of photon counting detector were analyzed. For the sake of realizing the linear measurement of the photon counting, the pump power was adjusted to change incident photon rate, and the experimental measurements were corrected. The results show that the measurement precision can be improved by measurement and correction. The maximum linear range rate of the incident photon could be expanded from 105/s to 107/s result from lower non-linear errors.
    HU You-bo, LI Jian-jun, XIA Mao-peng, GAO Dong-yang, ZHENG Xiao-bing. Measurement and Correction of Linearity of Photon Counters[J]. Acta Photonica Sinica, 2016, 45(6): 604001
    Download Citation