• Journal of Atmospheric and Environmental Optics
  • Vol. 10, Issue 5, 432 (2015)
Linan ZHENG*, Shunxing HU, Jian HUANG, Ke’e YUAN, and Shisheng SHAO
Author Affiliations
  • [in Chinese]
  • show less
    DOI: 10.3969/j.issn.1673-6141.2015.05.010 Cite this Article
    ZHENG Linan, HU Shunxing, HUANG Jian, YUAN Ke’e, SHAO Shisheng. An Improved Circuit Design for Atmospheric CO2 Balloonsonde Measurement[J]. Journal of Atmospheric and Environmental Optics, 2015, 10(5): 432 Copy Citation Text show less

    Abstract

    A method is presented for atmospheric CO2 balloonsonde measurement. Based on Beer-Lambert law and non-dispersed infrared(NDIR) technique, an experimental device is designed and developed. An infrared LED is used to produce suitable light source, and infrared LED detectors and a proper designed electronic circuit are used to transform light intensity into electrical signals in the experimental device. Calibration method for CO2 measurement is also presented. A comparison experiment is carried out for continuous 12 days on ground with EC9820 analyzer. Result shows that the trend of the measurement data of 12 days is consistent, and standard deviation is 9.835 ppm, which basically satisfies the precision demand for atmospheric CO2 measurement. The feasibility of CO2 measurement has been validated.
    ZHENG Linan, HU Shunxing, HUANG Jian, YUAN Ke’e, SHAO Shisheng. An Improved Circuit Design for Atmospheric CO2 Balloonsonde Measurement[J]. Journal of Atmospheric and Environmental Optics, 2015, 10(5): 432
    Download Citation