    Abstract

    In order to measure spatially and temporarily resolved laser-produced plasma X-ray spectra in 0.2-2 nm region, a novel two-channel elliptically-bent crystal spectrograph has been developed. Dispersive elements are LiF, PET, Mica, and KAP crystals, which cover Bragg angles in the range of 30-67.5 degrees. Eccentricity and focal distance of twin ellipses are 0.9586 and 1350 mm, respectively. Spatially resolved spectrum is photographically recorded with an X-ray film or X-CCD camera in one channel, and temporarily resolved one is photographically recorded with an X-ray streak camera in another channel, thus spatially and temporarily resolved spectra can be simultaneously obtained. Spectral images were acquired with X-CCD and PET in SHENGUANG-II laser facility, and experimental results show that the spectral resolution of the spectrograph is about 0.002 nm.
    Xiancai Xiong, Xianxin Zhong, Shali Xiao, Guohong Yang, Jie Gao. Elliptically-bent crystal spectrograph for X-ray diagnosis of laser-produced plasmas[J]. Chinese Optics Letters, 2004, 2(1): 0127