• Journal of Atmospheric and Environmental Optics
  • Vol. 4, Issue 3, 223 (2009)
Su-ning LI*, Ri-hong ZHU, Jian-xin LI, and Yan WANG
Author Affiliations
  • [in Chinese]
  • show less
    DOI: Cite this Article
    LI Su-ning, ZHU Ri-hong, LI Jian-xin, WANG Yan. Discussion about the Spectral Resolution of the Sagnac Interference Imaging Spectrometer Caused by Offset of the Sagnac Prism[J]. Journal of Atmospheric and Environmental Optics, 2009, 4(3): 223 Copy Citation Text show less

    Abstract

    As to improve the spectral resolution of the Sagnac interference imaging spectrometer, the mathematical model of the system is analyzed. The effect of the offset of the Sagnac prism on the spectral resolution of the system is discussed, and the concrete numerical value of the prism offset is calculated. The simulation experiment is carried out by importing the simulation spectral curve and assuming that it was error free in the instrument installation. The result shows that the spectral resolution can be 2 nm by adapting the calculated offset value, which has the guiding significance for the structure design of the system.
    LI Su-ning, ZHU Ri-hong, LI Jian-xin, WANG Yan. Discussion about the Spectral Resolution of the Sagnac Interference Imaging Spectrometer Caused by Offset of the Sagnac Prism[J]. Journal of Atmospheric and Environmental Optics, 2009, 4(3): 223
    Download Citation