• Journal of Atmospheric and Environmental Optics
  • Vol. 14, Issue 4, 296 (2019)
Mili BAI1、2、*, Jianjun LI1, Dongyang GAO1, Youbo HU1, Chenyu XIE1, Jing YAN1, and Xiaobing ZHENG1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3969/j.issn.1673-6141.2019.04.008 Cite this Article
    BAI Mili, LI Jianjun, GAO Dongyang, HU Youbo, XIE Chenyu, YAN Jing, ZHENG Xiaobing. Comparison Experiment of Two-Channel Self-Calibration Radiometer and Radiance Standard Detector[J]. Journal of Atmospheric and Environmental Optics, 2019, 14(4): 296 Copy Citation Text show less

    Abstract

    High-precision self-calibration radiation reference source based on spontaneous parametric down-conversion is the research hotspot of satellite remote sensing calibration. The principle of two-channel self-calibration radiometer was introduced, and a set of experimental device was established. The experimental verification was carried out by comparing with traditional radiometric standard detector. Monochromatic uniform radiance light source with 736.9 nm laser was used by integrating sphere, then the radiance at the outlet of integrating sphere was detected by two-channel self-calibration radiometer and radiance standard detector respectively. The measured absolute radiances results were 4.69×10-9 W/(sr·mm2) and 5.19×10-9 W/(sr·mm2), and the relative deviation was 9.63%. The feasibility of the principle of two-channel self-calibration radiometer was verified.
    BAI Mili, LI Jianjun, GAO Dongyang, HU Youbo, XIE Chenyu, YAN Jing, ZHENG Xiaobing. Comparison Experiment of Two-Channel Self-Calibration Radiometer and Radiance Standard Detector[J]. Journal of Atmospheric and Environmental Optics, 2019, 14(4): 296
    Download Citation