• Acta Photonica Sinica
  • Vol. 39, Issue 6, 1132 (2010)
MA Ying1、2、*, LIU Qun1、2, QIU Dan-gui1、2, SHI Ying-xia1、2, HOU Jian-Rong1、2, and YIN Wang-Bao1、2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: Cite this Article
    MA Ying, LIU Qun, QIU Dan-gui, SHI Ying-xia, HOU Jian-Rong, YIN Wang-Bao. Optical Measuring Concentration of DOP Aerosol[J]. Acta Photonica Sinica, 2010, 39(6): 1132 Copy Citation Text show less

    Abstract

    In order to ensure the work efficiency of High Efficiency Particulate Air Filter(HEPA),Dioctylphthalate method,which is widely used in the efficiency and leaking test of HEPA filtration system,is analyzed.The optical detect theory of Ddictylphthalate aerosol concentration is summarized.Based on Mie scattering theory,a mathmatical model is derived on relation between light-scattering intensity and concentration of aerosol when the special aerosol is generated.The optical measurement system based on the theory is designed and manufactured.The validity of the model is testified by the experiment.The experimental result shows that there is a good linear relation between light-scattering intensity and mass concentration of Dioctylphthalate aerosol.
    MA Ying, LIU Qun, QIU Dan-gui, SHI Ying-xia, HOU Jian-Rong, YIN Wang-Bao. Optical Measuring Concentration of DOP Aerosol[J]. Acta Photonica Sinica, 2010, 39(6): 1132
    Download Citation