• Chinese Optics Letters
  • Vol. 11, Issue s2, S21701 (2013)
Zhong Ren, Guodong Liu, and Zhen Huang
DOI: 10.3788/col201311.s21701 Cite this Article Set citation alerts
Zhong Ren, Guodong Liu, Zhen Huang. Investigation of glucose concentration measurement based on tunable pulsed laser induced photoacoustic technique[J]. Chinese Optics Letters, 2013, 11(s2): S21701 Copy Citation Text show less

Abstract

A tunable pulsed laser induced photoacoustic (PA) measurement set-up is established in the forward mode to monitor in vitro glucose concentration. A series of experiments are investigated to verify the feasibility of this set-up and scheme. Peak-to-peak values (PPVs) of several glucose aqueous solutions are recorded and averaged 512 times at each wavelength. Experimental results demonstrate that the time-resolved PA profile of glucose solutions has a good agreement with the PA theories. The characteristic wavelengths of glucose solution are determined via differential method. The root-mean-square error (RMSE) of predicted concentration reaches 3.15 mg/dl at the optimum wavelength of 1 510 nm via least square (LS) fitting algorithm.
Zhong Ren, Guodong Liu, Zhen Huang. Investigation of glucose concentration measurement based on tunable pulsed laser induced photoacoustic technique[J]. Chinese Optics Letters, 2013, 11(s2): S21701
Download Citation