• Acta Photonica Sinica
  • Vol. 39, Issue 3, 431 (2010)
JIN Qing-li*, YAN Li-fen, and WANG Zhen-guo
Author Affiliations
  • [in Chinese]
  • show less
    DOI: Cite this Article
    JIN Qing-li, YAN Li-fen, WANG Zhen-guo. Real-time High Sensitivity Variation of Solution Concentration Monitor Based on Fabry-Perot Interferometer[J]. Acta Photonica Sinica, 2010, 39(3): 431 Copy Citation Text show less

    Abstract

    A real-time solution concentration monitor based on Fabry-Perot interferometer is designed.The concentration variation measurement is realized by the F-P interferometer and optical fiber to transmit the image to the CCD.The relation between the variation of interference fringes of the F-P interferometer and the fluctuation of solution concentration is analyzed.The monitor system consists of He-Ne laser,optical fiber,planar F-P interferometer,arrayed CCD and PC is set up.Three different glycerin solution are used to verify the system design,and the experiment shows the system can provide a resolution on the order of 10-4 concentration variation.
    JIN Qing-li, YAN Li-fen, WANG Zhen-guo. Real-time High Sensitivity Variation of Solution Concentration Monitor Based on Fabry-Perot Interferometer[J]. Acta Photonica Sinica, 2010, 39(3): 431
    Download Citation