• High Power Laser and Particle Beams
  • Vol. 33, Issue 4, 046002 (2021)
Yikun Qian1、2, Peng Feng2, Yixin Liu1, Yadi Zhu1, Yuchen Huang1、2, Song Zhang1、2, Peng He2, Biao Wei2, and Benjiang Mao1、*
Author Affiliations
  • 1China Academy of Engineering Physics, Mianyang 621900, China
  • 2College of Optoelectronics Engineering, Chongqing University, Chongqing 400044, China
  • show less
    DOI: 10.11884/HPLPB202133.200280 Cite this Article
    Yikun Qian, Peng Feng, Yixin Liu, Yadi Zhu, Yuchen Huang, Song Zhang, Peng He, Biao Wei, Benjiang Mao. Personal dosimeter calibration based on minitype reference radiation[J]. High Power Laser and Particle Beams, 2021, 33(4): 046002 Copy Citation Text show less
    [in Chinese]
    Fig. 1. [in Chinese]
    [in Chinese]
    Fig. 2. [in Chinese]
    [in Chinese]
    Fig. 3. [in Chinese]
    [in Chinese]
    Fig. 4. [in Chinese]
    [in Chinese]
    Fig. 5. [in Chinese]
    [in Chinese]
    Fig. 6. [in Chinese]
    [in Chinese]
    Fig. 7. [in Chinese]
    thickness of stage/mmdose rate at the point of test/(mSv/h)impact rate/%error/%
    51.3831.140.11
    101.3952.000.11
    151.4042.700.11
    201.4123.270.11
    251.4193.780.11
    301.4244.200.11
    351.4304.570.09
    401.4344.900.09
    Table 1. [in Chinese]
    size of dosimeter/(mm×mm×mm)dose rate at the point of test/(mSv/h)impact rate/%error/%
    10×3×81.4271.620.03
    8×3×81.4191.080.04
    8×3×61.4110.460.04
    8×3×41.397−0.470.05
    8×2×41.4140.700.05
    8×1×41.4201.110.05
    Table 2. [in Chinese]
    number of dosimeter/mmdose rate at the point of test/(mSv/h)impact rate/%error/%
    11.4000.100.10
    21.4010.150.09
    41.4020.240.10
    81.4070.560.11
    Table 3. [in Chinese]
    uncertainty factorreference conditionerror value/%uncertainty component/%
    MRR box60 cm×60 cm×60 cm0.080.05
    thickness change of stage20 mm3.271.89
    size change of dosimeters10 mm×3 mm×8 mm1.620.94
    number change of dosimeters80.560.32
    combined standard uncertainty3.20