    Qing-bo LI, Yuan WEI, Hou-xin CUI, Hao FENG, Jia-ye LANG. Quantitative Analysis of TOC in Water Quality Based on UV-Vis Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 376 Copy Citation Text show less
    Absorbance spectra of TOC samples with different concentrations
    Fig. 1. Absorbance spectra of TOC samples with different concentrations
    PRESS curve of the predicted concentration of the test set samples using different principal component numbers
    Fig. 2. PRESS curve of the predicted concentration of the test set samples using different principal component numbers
    Relative error of sample predicted concentration in test set using the optimal principal component number
    Fig. 3. Relative error of sample predicted concentration in test set using the optimal principal component number
    Spectral comparison of TOC actual concentration of 28 and 31 mg·L-1 on different experimental days
    Fig. 4. Spectral comparison of TOC actual concentration of 28 and 31 mg·L-1 on different experimental days
    Relative error of predicted TOC concentration in experiment 2
    Fig. 5. Relative error of predicted TOC concentration in experiment 2
    实际浓度/
    (mg·L-1)
    预测浓度/
    (mg·L-1)
    相对误差/
    %
    37.0037.451.21
    49.0049.521.07
    61.0061.570.93
    76.0075.950.06
    88.0088.630.72
    109.00106.462.33
    127.00127.590.46
    145.00146.551.07
    Table 1. TOC concentration prediction results of validation experiment with different instrument states in the first experimental day
    实验类型预测浓度最大
    相对误差/%
    MAPE/%RMSEP
    实验17.193.823.75
    实验215.913.752.99
    实验36.483.433.91
    实验42.330.981.14
    Table 2. TOC concentration prediction results of cross experiment with different instrument status on different experimental days
    Qing-bo LI, Yuan WEI, Hou-xin CUI, Hao FENG, Jia-ye LANG. Quantitative Analysis of TOC in Water Quality Based on UV-Vis Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 376