• Infrared and Laser Engineering
  • Vol. 45, Issue 7, 705005 (2016)
Zhao Kun1、*, Shi Xueshun1、2, Liu Changming1, Liu Yulong1, Chen Haidong1, Liu Hongbo1, Chen Kunfeng1, and Li Ligong1、2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    DOI: 10.3788/irla201645.0705005 Cite this Article
    Zhao Kun, Shi Xueshun, Liu Changming, Liu Yulong, Chen Haidong, Liu Hongbo, Chen Kunfeng, Li Ligong. Laser source used in accurate measurement of mid-infrared absolute spectral responsivity of detectors[J]. Infrared and Laser Engineering, 2016, 45(7): 705005 Copy Citation Text show less

    Abstract

    In order to measure the absolute spectral responsivity of mid-infrared detectors with high accuracy, the improvement of power stability and beam quality of mid-infrared laser was theoretically and experimentally demonstrated. The power stability of mid-infrared laser was improved to be better than 0.1% by using acousto-optic modulator and feedback-control electronics. Based on propogation properties of Gaussion beam, the parameters were calculated and experimental setup of spatial filter was constructed. Owing to the spatial filter, the beam quality of mid-infrared laser was improved. A stable and reliable laser source is provided for the measurement of absolute spectral responsivity of mid-infrared detectors.
    Zhao Kun, Shi Xueshun, Liu Changming, Liu Yulong, Chen Haidong, Liu Hongbo, Chen Kunfeng, Li Ligong. Laser source used in accurate measurement of mid-infrared absolute spectral responsivity of detectors[J]. Infrared and Laser Engineering, 2016, 45(7): 705005
    Download Citation