Abstract
1 Introduction
Volumetric tissue imaging based on multimodal microscopy techniques [e.g., confocal microscopy, multiphoton microscopy, harmonic generation microscopy, and coherent Raman scattering (CRS) microscopy]1
In this work, we present a unique phase-modulated stimulated Raman scattering tomography (PM-SRST) technique for rapid 3D chemical imaging in living cells and tissue. To achieve the mechanical scan-free sectioning in PM-SRST, we apply a spatial light modulator (SLM) to manipulate the axial position of the focused Gaussian Stokes beam along the needle Bessel pump beam in the sample for SRST imaging. PM-SRST facilitates the direct acquisition of 3D sample information in the spatial domain, without necessitating postprocessing procedures. We have derived and analyzed the unique Bessel beam-based PM-SRST imaging method theoretically and also built the PM-SRST imaging system to experimentally demonstrate its capability for rapid label-free volumetric chemical imaging on a variety of samples [e.g., Brownian motion of polymer beads in water, diffusion and uptake processes of deuterium oxide () in the plant root, and breast cancer cell response to acetic acid]. We have also demonstrated the utility of the PM-SRST technique for improving both the lateral resolution and imaging depth of SRS 3D imaging of biotissues.
2 Materials and Methods
2.1 Working Principle of PM-SRST
In SRS imaging, the sample is illuminated by the collinear pump and Stokes beams under a tight focus through a high numerical aperture (NA) objective lens. The SRS signal generated is proportional to the product of the pump and Stokes beam intensities.4,5 In PM-SRST, the pump light is converted into a Bessel beam by using an axicon and is focused onto the sample, in which the intensity distribution in the sample space is as follows:15
Sign up for Advanced Photonics TOC. Get the latest issue of Advanced Photonics delivered right to you!Sign up now
The Gaussian Stokes beam is focused onto the sample as an Airy spot under the high NA objective lens, in which the intensity distribution is , where is the on-axis intensity of the incident Gaussian Stokes beam and is the waist of the Stokes beam. Hence, the SRS intensity at position is13
To control the axial position of the focused Stokes spot in the sample for depth-resolved SRS imaging, we change the divergence angle of the incident Stokes beam at the Fourier plane of the system through changing the phase patterns of the Stokes beam via an SLM. We can calculate the optical path difference (OPD) between the incident plane wave and the spherical wavefront to determine the axial position of the focal spot away from the focal plane of the lens. According to the Gaussian thin lens formula,16 the relationship between the axial position of the incident point source and the focus is as follows:
By substituting Eq. (5) into Eq. (2), we obtain the relationship between the SRS intensity on sample and as follows:
2.2 Development of PM-SRST System
Figure 1(a) illustrates the schematic of the PM-SRST system developed for 3D chemical imaging. A 100-fs laser (InSight, Spectra Physics) operating at 80 MHz repetition rates provides dual color outputs: a fixed 1041 nm beam (Stokes), and a tunable (680 to 1300 nm) beam (pump) for generating an SRS signal. The Stokes beam is modulated at 20 MHz by an electro-optic modulator (EOM) and then projected onto an SLM (HCA512-780-940, Meadowlark Optics). After modulating the phase patterns of the Stokes beam by the SLM, the +1st-order of the diffracted Stokes light is selected for SRS imaging, while the other orders of the diffracted beams are blocked by an aperture. For the pump path, the beam is converted into a Bessel beam after passing through a 0.5 deg axicon (AX2505, Thorlabs). Then the Bessel pump and Gaussian Stokes beams are combined by a dichroic mirror (DMLP900R, Thorlabs) and coupled into a laser-scanning microscope (MPM-4R, Thorlabs) with a galvo-scanner (GVSM002, Thorlabs), and then focused onto the sample with a high NA objective lens (CFI75 APOCHROMAT 25XW MP 1300, NA = 1.1, Nikon). A high NA condenser (CC Achromat/Aplanat, NA = 1.4, Nikon) is used for collecting forward SRS emission from the sample. In the detection path, the Stokes beam is spectrally removed by using a bandpass filter (795/75, Semrock), while the modulated pump beam is detected by a large-area photodiode (FDS1010, Thorlabs) with a lock-in amplifier for demodulating the stimulated Raman loss signal from the pump beam for SRS 3D imaging.
Figure 1.(a) Schematic of PM-SRST system for 3D chemical imaging. EOM, electro-optical modulator; SLM, spatial light modulator; Ape, aperture; DM, dichroic mirror; GM, galvo mirror; SL, scan lens; TL, tube lens; F, filter; LIA, lock-in amplifier; PD, photodiode. (b) Schematic of the working principle of PM-SRST. The pump beam is converted into a Bessel beam, and the Stokes beam is focused into the sample. The depth-resolved SRS image is achieved by manipulating the axial position of the focused Gaussian Stokes beam and the overlapping with the needle Bessel pump beam at different depths by SLM. (c) Raw SRS images of
Figure 1(b) shows the schematic of the working principle of the PM-SRST technique. To achieve the depth-resolved SRS 3D imaging, the pump beam is a needle Bessel beam in the sample. The axial position of the Gaussian Stokes spot is manipulated along the needle Bessel pump beam by controlling the divergent angles of the incident Gaussian Stokes beam with the SLM before entering the objective lens. The modulated phase patterns corresponding to specific depths [Eq. (5)] are generated [Fig. S1(a) in the Supplementary Material] and varied on the SLM for 3D chemical imaging. For example, the raw SRS images of polystyrene (PS) beads are acquired at , , and , respectively [Fig. 1(c)], illustrating the depth-sectioning capability of PM-SRST experimentally.
2.3 Sample Preparation
The PS beads (17136, Polysciences) embedded in 2% agarose gel phantoms are used for 3D imaging (). The PS bead suspension (17135, Polysciences, diluted 10 times with water) is dropped onto the microscope slides with a single well (MS15C1, Thorlabs) for SRS monitoring of Brownian motion. The 500 nm fluorescent beads (19507, Polysciences) are fixed on the microscope slides for measuring the intensity distributions of the focused Gaussian Stokes beam and the needle Bessel–Stokes beam in the and planes. The 500 nm PS beads (07307, Polysciences) are fixed on microscope slides for the assessment of the point spread function (PSF) (i.e., lateral resolution) of the PM-SRST technique.
For imaging the plant’s root, the seed of choy sum is cultured in the soil containing okara-based hydrogels17 under the conditions of the temperature of (light/dark) with RGB LED lights [ under 12 h light on/12 h light-off illumination] for about half of a month, and the intact root of the plant is cut by surgical blades and fixed on the microscope slide with nail polish. Then (151882-10X0.6ML, Sigma) is dropped onto the root for SRST monitoring of the diffusion and uptake processes of in the plant root.
The breast cancer cell line (MCF-7) is cultured in the medium containing a high-glucose Dulbecco’s modified eagle medium (DMEM), 1% penicillin-streptomycin, and 10% FBS. Cells are subcultured and plated on a bottom glass (MatTek Co., Ashland, MA) for three days in the incubator. Before SRST imaging, the cells are washed twice with PBS to remove excess media. About 100% acetic acid is dropped into 0.5 mL PBS to treat the cells.
The fresh porcine skin and brain (purchased from the local supermarket) tissue slides with tissue thicknesses of and , respectively, are prepared by using surgical blades and sandwiched in between the coverslips for SRST imaging.
2.4 SRS Imaging Parameters
The time constant of in the lock-in amplifier is used for SRS imaging. A 2D raster scan rate of 51.2 Hz (pixel dwell time of , pixel numbers of ) is used for monitoring the Brownian motion of PS beads in water. For other larger image sizes (pixel numbers of ), a typical 2D raster scan rate of 3.5 Hz (pixel dwell time of ) is employed in PM-SRST and conventional SRS imaging. The wavelengths of the pump and Stokes beams are 800 and 1041 nm, respectively.
2.5 3D Image Acquisition and Reconstruction
For rapid SRST 3D imaging, the SLM is triggered and synchronized by the microscope scanning module. The phase patterns generated (Fig. S1 in the Supplementary Material) are projected on the SLM (operating at 60 Hz refresh rate) for acquiring the 2D images at the corresponding depths. A series of 2D images are used to reconstruct 3D volume with ImageJ. For comparison purposes, the conventional point-scan SRS (C-SRS) 3D imaging is also acquired by using a motorized vertical translation stage (switching rate of 10 Hz, KVS30, Thorlabs) to move the sample step-by-step along the axis.
2.6 Estimations of Spatial Resolution and Light Penetration Depth of PM-SRST
The lateral resolutions of the Gaussian pump and Stokes beam are calculated by and , respectively, where the and are the wavelengths of the pump and Stokes beams, respectively. NA is the numerical aperture. The PSF (main lobe) of the Bessel pump beam is calculated by 18. The lateral resolution of the SRS imaging technique is determined by the superposition of the pump and Stokes beams. Hence, the lateral resolutions of PM-SRST and C-SRS imaging are calculated by and , respectively14. With the use of , and NA = 0.45 in this study, the lateral resolution of PM-SRST is calculated to be , while the lateral resolution of C-SRS is . In theory, PM-SRST provides -fold improvement in lateral resolution as compared to the C-SRS imaging. The axial resolution of the Gaussian pump and Stokes beam can be calculated by and , respectively. Therefore, the axial PSF of C-SRS is calculated to be using (NA = 0.45), while the axial resolution of PM-SRST is equivalent to the axial PSF of the Gaussian Stokes beam, which is . In theory, the axial resolution of C-SRS is -fold better than that of PM-SRST. Note that the Gaussian function is employed to fit the intensity distribution of 500 nm PS beads (Fig. 2), in which the full width at half-maximum (FWHM) of the fit curve stands for the resolution of PM-SRST and C-SRS techniques.
Figure 2.(a) and (c) Lateral SRS images (
The exponential decay function is used for the calculation of penetration depth, where and are the average SRS intensities at the surface and depth , respectively. is the penetration depth, when the SRS intensity is dropped by of the incident light intensity .
2.7 Method for the Diffusion Rate Estimation
For estimating the concentration variations of with time in the plant root [Fig. 6(g)], the water diffusion and uptake model, is used19, where is the water concentration in the root and is the one at the root surface. The is the diffusion permeability across the membrane of the cortex cell in the apoplast of the root. is defined as the thickness of the cortex, corresponding to the step size () in PM-SRST.
2.8 Definition of Average SRS Intensity, Diameter, and Volume Size of Nucleus in Cancer Cells
To assess the biochemical variations in cancer cells subjected to acetic acid stimulant, the pixel intensity averaging is performed at each depth to derive the average SRS intensity of the cytoplasm within a single cell by excluding the nucleus region. Similarly, the average pixel intensity of the nucleus at each depth is used for calculating the average SRS intensity of the nucleus. The intensity of nucleus is fit with a Gaussian function, and the FWHM of the fit curve is used to estimate the diameter of nucleus. The volume size of nucleus is calculated by multiplying the nucleus area at each depth by the axial step size used.
3 Results
3.1 PM-SRST System Characterization
Figures 2(a)–2(d) show the comparison of the lateral and axial resolutions between the PM-SRST and C-SRS imaging on a 500 nm PS bead. The corresponding lateral resolutions for PM-SRST and C-SRS imaging are 0.65 and , respectively [Fig. 2(e)], giving a -fold improvement in lateral resolution achieved by PM-SRST. The axial resolution of PM-SRST is [Fig. 2(f)], which is slightly lower than that of C-SRS imaging (). The experimental measurement result above is close to the theoretical values (Sec. 2.6). The slight discrepancies in spatial resolutions between the experiment and theoretical calculations may be due to the slightly lower NA for the pump and Stokes beams used in PM-SRST experiments.
We also compare the light-penetration depth of PM-SRST with C-SRS imaging by measuring SRS images of the PS beads in agarose gel phantoms [Fig. 3(a)]. The -scanning-free PM-SRST imaging shows almost the same PS beads 3D distributions in the gel phantom as those in conventional mechanical point-scanning SRS imaging, but the SRS signal from the deeper bead areas in PM-SRST is much stronger than that in C-SRS (e.g., the bead pointed by the yellow arrow in the deeper area of the tissue phantom is much brighter in PM-SRST compared to that in C-SRS). To quantitatively analyze the imaging penetration depth by PM-SRST, Fig. 3(b) shows the normalized SRS intensities of the beads along the depth in PM-SRST as compared to C-SRS imaging. The plots are fit by the exponential decay function (refer to Sec. 2.6). The penetration depth [95% confidence interval: (, )] is obtained for PM-SRST, while [95% confidence interval: (, )] for C-SRS. The PM-SRST technique provides ∼twofold improvement in penetration depth in polymer beads phantom as compared to C-SRS imaging, affirming the potential of PM-SRST for deeper tissue 3D imaging.
Figure 3.(a)
To demonstrate the rapid 3D imaging capabilities of PM-SRST, we monitor the 3D Brownian motion of PS beads in water. Figure 3(c) displays a series of snapshot SRS 3D images of the moving beads in water captured at the 3D volume rate of 8.5 Hz (in a 118 ms time interval) (see Video 1 and Fig. S1 in Supplementary Material for details) without the image blurring of the beads traced. We also measure the 3D Brownian motion of of the bead, as indicated with a solid yellow arrow in Fig. 3(c), which is in agreement with the theoretical prediction of (with a time of 0.118 s and temperature of 293 K calculated with Stokes–Einstein equation 20), substantiating the potential of PM-SRST for 3D dynamic chemical imaging.
3.2 Monitoring of Diffusion and Uptake Processes of
We have applied this rapid PM-SRST technique to observe the 3D diffusion and uptake processes of in the fresh plant root. Figure 4(a) presents the 3D dynamic process of ( of the O-D chemical bond in the silent region) in the root captured by the PM-SRST within 5.84 s (Video 2). For the comparison purpose, Fig. 4(b) shows the SRS image () of the same tissue volume by PM-SRST in the absence of as a control. To better visualize the spatial distribution of within the plant root, an SRS 3D image of the root at is displayed in Fig. 4(c) [as exemplified in Fig. 4(d)]. Figure 4(e) shows the linear relationship between the SRS intensity and the concentration of for determining the distribution of in the plant root [Fig. 4(i)]. Additionally, Fig. 4(f) depicts the variation of SRS intensity along the axis for the intensity and concentration calibration of in Figs. 4(g) and 4(i). The SRS intensities of at each depth in the root are analyzed as shown in Fig. 4(g), showing an exponential increase over time. The plots fit with the water diffusion and uptake equation (refer to Sec. 2.7) in Fig. 4(g), and the diffusion permeabilities of (1.6 to ) are calculated at different depths of the root [Fig. 4(h)]. The results reveal that the deeper regions of the plant’s root exhibit a more rapid attainment of equilibrium concentration for . Furthermore, we estimate the concentration change of over time in the plant root [Fig. 4(i)]. Notably, the variations of the concentration versus the plant root’s depth at and exhibit a consistent trend, suggesting the rapid establishment of a dynamic equilibrium for diffusion in the plant roots.
Figure 4.(a) 3D diffusion dynamics of
3.3 Visualization of Biochemical Variations of Live Breast Cancer Cells Subjected to Acetic Acid by PM-SRST
We have also utilized PM-SRST to visualize the rapid biochemical variation of breast cancer cells (MCF-7) subjected to acetic acid stimulants. Figure 5(a) shows the snapshot 3D time-lapse SRS images () of the cells captured by PM-SRST (Video 3). An increasing trend in SRS intensity within the nucleus is observed (indicated by a red solid arrow). To gain deeper insights into this phenomenon, we investigate the temporal evolution of nucleus volume versus volume-averaging SRS intensity, as depicted in Fig. 5(c). The increase in SRS intensity within the nucleus observed coincides with a concurrent reduction in nucleus volume size, indicating a direct correlation between the enhanced SRS signal and the nucleus shrinkage in response to the acetic acid stimuli within seconds.
Figure 5.(a) 3D SRS dynamics of MCF-7 cells (
We also analyze the SRS intensity variation in the cytoplasm over time [Fig. 5(d)]. Initially, the SRS signals show an increasing trend but subsequently decline at each depth within the cytoplasm. These observations can be attributed to the localized concentration increase of biochemicals resulting from the rapid cell contraction and subsequent degradation of lipids/proteins inside the cytoplasm in cells.
To further investigate the biochemical variations in the cells, Fig. 5(b) depicts the SRS 3D dynamic process of the cells at captured by PM-SRST (Video 4). Figure 5(e) shows the change in SRS intensity over time inside the cytoplasm. We find that the SRS intensity at consistently increases during the stimulus, probably due to the localized intensity enhancement [indicated by the black solid arrows in Fig. 5(b)] caused by the cell shrinkage following exposure to acetic acid. Consequently, by comparing the SRS intensity variations at (associated with symmetric stretching of lipids) and ( stretching of lipids and proteins) in cells, we can infer the degradation of proteins within the cytoplasm of live cells during acetic acid stimuli.
3.4 Assessment of Deeper Penetration Capability of PM-SRST in Biotissues
We have further validated the capability of PM-SRST technique to improve the imaging depth in highly scattering biotissues (e.g., porcine skin and brain tissue) compared to C-SRS imaging. Figs. 6(a) and 6(b) present the SRS 3D images of porcine skin tissue captured by PM-SRST and C-SRS, confirming a higher SRS intensity observed from deeper regions of the tissue with the PM-SRST technique. To better visualize the improvement in penetration depth, Figs. 6(c) and 6(d) display the plane of the 3D images in Figs. 6(a) and 6(b), showing the enhanced imaging depth achieved by PM-SRST in comparison with C-SRS. Figure 6(e) shows the measured average SRS intensities at different tissue depths that are normalized to the average SRS intensity (all pixel intensities) at the tissue surface (). The fitting curves using the exponential decay function (refer to Sec. 2.6) indicate that the penetration depth in porcine skin by PM-SRST is , giving ∼twofold improvement in imaging depth compared to C-SRS imaging (). We have also compared the imaging depths between PM-SRST and C-SRS imaging in even stronger scattering porcine brain tissue.21Figures 6(f) and 6(g) present the 3D images of brain tissue along with plane image [Figs. 6(h) and 6(i)], confirming the significant improvement in penetration depth achieved by PM-SRST. The SRS intensity profiles along the direction in Fig. 6(j) show that the penetration depth for PM-SRST in brain tissue is , giving rise to a 2.3-fold improvement in imaging depth as compared to C-SRS imaging (). Thus, PM-SRST exhibits the enhanced penetration capability for achieving deeper 3D chemical imaging in biological tissues. The slight variations in the degree of improvement in imaging depths among different tissue samples may be attributed to the varying scattering effects encountered in different types of tissues [e.g., porcine skin (scattering coefficients for 800 nm and for 1041 nm) and brain tissue ( for 800 nm and for 1041 nm)]21.
Figure 6.(a) and (b) SRS 3D images (
4 Discussion
Conventional SRS 3D imaging coupled with mechanical scanning encounters the challenges in inertia artifacts and slow imaging speed in tissue and cells.12 The strong scattering effect in turbid media (e.g., tissue) also dramatically attenuates the light-propagating intensity in deeper tissue regions, limiting the imaging depth.22 To tackle these issues, we have developed a unique PM-SRST technique that enables rapid 3D chemical imaging by electronically tuning the superpositions of the focused Gaussian–Stokes beam along the needle Bessel pump beam without the need for mechanical scanning. The fast 3D imaging capability is achieved by rapidly controlling the axial positioning of the Stokes beam with a phase modulation device. We use the SLM as a phase modulator to evaluate the imaging speed of PM-SRST by monitoring the 3D Brownian motion of PS beads in water [Fig. 3(c)], in which the 3D volume rate of 8.5 Hz [i.e., 118 ms time interval, six phase patterns, Fig. S1(a) in the Supplementary Material] can be achieved. Compared with the deformable mirror-based volumetric SRS imaging,23 PM-SRST shows a remarkable superiority with millisecond time intervals, confirming the advantages of electronically-controlled phase modulations for improving depth-resolved SRS 3D imaging speed. To acquire more sectioning images along the axis, a faster phase modulation device could be employed for PM-SRST. Compared to an electrically tunable lens ( frame rate24) or a tunable acoustic gradient index of refraction lens (constrained dwelling capacity25) with device-induced spherical and high-order aberration,26 the microelectromechanical system (MEMS)-based SLM (kilohertz frame rate)24,27 or digital micromirror device (DMD, 4.2 kHz28) could be used to further boost the -sectioning speed of the PM-SRST technique for video-rate SRS 3D imaging of tissue and cells. PM-SRST also provides a deeper light-penetration depth for 3D chemical imaging in tissue compared to C-SRS imaging. We have experimentally demonstrated a twofold improvement in imaging depth in PS beads phantom by PM-SRST [Figs. 3(a) and 3(b)] compared to C-SRS imaging. To elucidate the reason for this improvement in penetration depth within a highly scattering medium, we estimate the scattering coefficients for the pump beam ( at 800 nm) and Stokes beam ( at 1041 nm) based on the Mie scattering model29 [the density of the beads used in the phantom is in SRS imaging; see Fig. 3(a)]. Our results indicate a lower scattering for the longer wavelength (1041 nm) in the beads phantoms. As a result, the penetration depth of C-SRS imaging is predominantly constrained by the higher scattering effect of the shorter wavelength pump beam (800 nm). In PM-SRST, we convert the Gaussian pump beam into a nondiffracting Bessel beam, which has been proven to be a relatively good scattering-resistant property with remarkable resilience to scattering effects in turbid media.18,30 Additionally, the utilization of a longer wavelength Stokes beam within the NIR-II window (1000 to 1700 nm) further enhances the capability for deeper tissue imaging compared to the NIR-I window (650 to 950 nm)31,32. Thus, by combining the advantages of the scattering-resilient Bessel pump beam (800 nm) and the longer wavelength of Stokes beam (1041 nm), PM-SRST successfully boosted the SRS 3D imaging depth in biotissues [e.g., porcine skin and brain tissue (Fig. 6)]. Compared with the conventional adaptive optics (AO) technique, which is widely used for deep tissue imaging by monitoring the aberration of the light wavefront in the sample with a wavefront sensor and subsequently correcting the beam distortion with DM,33 PM-SRST achieves an improved imaging depth by using the unique scattering-resilient Bessel pump beam together with NIR-II Stokes beam, enabling the rapid 3D deeper imaging in tissue without the time delay associated with the wavefront detection and compensation in AO. On the other hand, the Bessel pump beam used in PM-SRST provides a 1.6-fold narrower main lobe than that of a Gaussian pump beam (refer to Sec. 2.6). Hence, the PM-SRST technique yields a -fold enhancement in lateral resolution compared to C-SRS [Fig. 2(e)]. However, the modulation of the Gaussian Stokes beam in PM-SRST operates under the paraxial condition [Eq. (5)], which may limit the maximum utilization of the tightly focused Stokes beam (e.g., NA 1.1, nonparaxial condition) in PM-SRST. As such, the lateral resolution of PM-SRST at the nonparaxial condition would be constrained to a certain extent. Due to the extended length of the Bessel beam compared to the Gaussian spot, the overlapped portion of the pump and Stokes beams in PM-SRST along the -axis direction is a bit larger than that of the C-SRS. Consequently, the axial resolution of PM-SRST is slightly compromised in comparison to C-SRS [Fig. 2(f)]. To mitigate this effect for enhancing the spatial resolution, the deconvolution algorithms (e.g., Richardson–Lucy algorithm34) can be incorporated into the PM-SRST technique, offering the potential for rapid superresolution imaging in deep tissue.
The PM-SRST technique presented in this work is fundamentally different from our previously reported Bessel beam-based OBT SRS 3D imaging technique14 in terms of its working principles and imaging performance. (i) PM-SRST technique works by electronically tuning the focused Gaussian Stokes beam along the needle Bessel pump beam with the SLM, facilitating the direct acquisition of 3D sample information in the spatial or time domain without necessitating the postprocessing procedures. In contrast, OBT operates by encoding the 3D information with Bessel beam beatings after hundreds of varying spatial frequency interval scans () in the frequency domain, and the volumetric information is reconstructed through the inverse Fourier transform. (ii) Our two-photon fluorescence (TPF) experiments (refer to Fig. S2 in the Supplementary Material) show that the tightly focused Gaussian Stokes beam used in PM-SRST is -fold stronger in local laser intensity as compared to the Bessel Stokes beam used in OBT-SRST. Hence, the PM-SRST technique using the Bessel pump beam coupled with the tightly focused Gaussian Stokes excitation scheme can significantly enhance the signal-to-noise ratio (SNR) over OBT-based SRST, achieving a much faster SRS 3D imaging speed in cells and tissue. (iii) Given the similar axial resolutions (, NA = 0.45 for PM-SRST; , for OBT), PM-SRST facilitates rapid 3D imaging within 10 s [33 scans of 2D images, depth in Figs. 6(a) and 6(b)], and the PM-SRST 3D imaging can be completed even down to 0.118 s (8.5 Hz, laser power of 48 mW), as shown in Fig. 3(c). In contrast, of OBT-SRS 3D imaging time (141 scans of 2D images, maximum depth, scanning rates of 12 s, laser power of 100 mW) is required for reconstructing a single 3D volume in OBT 3D imaging.14 Therefore, PM-SRST significantly boosts the 3D imaging speed with high SNRs and imaging depths for monitoring the dynamic functional and metabolic processes in tissue and cells with a lower incident laser power compared to the Bessel beam-based OBT SRS 3D imaging.
In summary, we have developed a unique PM-SRST technique for rapid label-free 3D chemical imaging in deeper tissue and cells without the need for mechanical scanning. We demonstrate the ability of PM-SRST technique for rapid dynamic/functional, and morphological 3D imaging in a variety of applications (e.g., Brownian motion monitoring, instant response of breast cancer cells to acetic acid stimuli, diffusion and permeation of in plant roots). One notes that the -scanning-free optical sectioning property with the phase control conceived in PM-SRST is universal, which can be easily extended to other imaging modalities for advancing 3D microscopy imaging in biological and biomedical systems. For instance, the current PM-SRST system can be readily adapted for CARS tomography by substituting the photodiode with a photomultiplier tube to capture 3D CARS signals. By utilizing either the pump or Stokes beam alone, the PM-SRST technique can be simplified to facilitate second-/third-harmonic generation tomography, multiphoton tomography, or fluorescence tomography, etc. We anticipate that the PM-SRST technique holds great promise to serve as a rapid -scanning-free 3D chemical imaging tool for a wide range of label-free biomolecular imaging applications in biology, biochemistry, and biomedicine.
Weiqi Wang received his BS degree in optical information science and technology from Ocean University of China in 2016 and his MSc degree in instrument science and technology from Tsinghua University, China in 2019. He is currently pursuing his PhD in Optical Bioimaging Laboratory under the supervision of Prof. Zhiwei Huang in the Department of Biomedical Engineering at the National University of Singapore. His research is mainly focused on stimulated Raman scattering tomography development and biological and biomedical applications.
Zhiwei Huang received his PhD from Nanyang Technological University, Singapore in 1999; he is the Director of Optical Bioimaging Laboratory in the Department of Biomedical Engineering, College of Design and Engineering at the National University of Singapore. His major research areas are in the fields of biomedical optics, microscopy, Raman spectroscopy and imaging, particularly centering on the development of superresolution microscopy and nonlinear optical microscopy imaging techniques (e.g., coherent Raman scattering microscopy, multiphoton microscopy) and their applications in biomolecular imaging, as well as the development of novel fiber-optic Raman spectroscopy and endoscopic imaging, enabling early diagnosis and detection of epithelial precancer and cancer at endoscopy. He pioneered in Raman endoscopy and label-free superresolution bioimaging technologies and published over 120 peer-reviewed articles in leading journals (e.g., Nature Photonics, Gastroenterology). He has filed over 20 U.S. patents, with 10 licensed for commercialization. His IMDX technique invented was ranked No. 1 among the top 10 medical devices listed in Medica, Germany. He is an elective Fellow of SPIE, the international society for optics and photonics.