Abstract
Introduction
The terahertz band (0.1-10 THz) have numerous characteristics,such as high resolution,penetrability[
KIDs is a type of superconducting terahertz detector with high sensitivity,which is sensitive in the ultraviolet to near-infrared range. Compared to semiconductor detectors,the advantages of KIDs in their operating wavelength range include no false counts (readout noise,dark current,and cosmic rays),intrinsic spectral resolution,high speed,and radiation resistance. Additionally,KIDs are more easily scalable into large detector arrays compared to other superconducting detectors for terahertz wavelengths.
After the fabrication of KIDs chips,corresponding testing systems are required to measure various parameters to characterize the performance of the detectors. Based on the working principle of KIDs,parameters such as S21,Power Spectral Density (PSD) of noise,and responsivity need to be measured. By processing the test parameters data using software algorithms,the Noise Equivalent Power (NEP) of KIDs can be obtained. NEP is a critical indicator of the detector's sensitivity,and high sensitivity is one of the most important performance metrics for detectors,because of detectors with high sensitivity can detect even weaker signals. For detecting distant celestial objects,the emitted signal power from these targets is extremely weak. Only high-sensitivity detectors like KIDs can detect such faint signals,meeting the requirements of astronomical observations. Therefore,it is crucial to measure the various performance parameters of KIDs before integrating them into astronomical detection systems.
The homodyne mixing technique is employed in the circuit system to characterize various aspects of KIDs' performance,and the performance of the homodyne mixing system significantly influences the characterization of the detector's performance. Designing a high-performance homodyne mixing system will enhance the accuracy of characterizing the detector's performance,thus contributing to improving the overall design of the detection system. Moreover,the easiness of operation and integration improvement of the homodyne mixing system will facilitate rapid testing and design work for researchers and designers,thereby enhancing the efficiency of KIDs' development. Due to the high sensitivity of KIDs itself,it is susceptible to external signal interference,and the homodyne mixing system inevitably introduces some interference. Designing a high-performance,convenient,highly integrated,and anti-interference homodyne mixing system is significant for the performance characterization,design,and testing of KIDs.
1 Principles of KIDs and Homodyne Mixing System
1.1 KIDs
KIDs The detection principle of KIDs is shown in
Figure 1.Principle of KIDs’ detection
By characterizing the changes in the transmission properties of the superconducting transmission line coupled to the superconducting resonator,the variations in resonant frequency and phase can be obtained. Consequently,the energy of the incident photons can be derived,and signal detection can be accomplished through the conversion of physical property relationships. A typical physical representation of KIDs is shown in
Figure 2.Physical image of typical KIDs
1.2 homodyne mixing system
The typical schematic diagram of the system using the homodyne mixing technique to characterize the performance of superconducting KIDs is shown in
Figure 3.Principle of homodyne mixing system structure for typical characterization of superconducting KIDs performance
The LO signal and the RF input are used as the input signals for the IQ-mixer. The two ADC signals(I(t) and Q(t) obtained from the time-domain signals fed into sampling ports of the ADC through low-pass filters and impedance transformers. The two sampled time-domain signals are theoretically orthogonal signals and correspond to the set frequency f of the signal generator. After applying Fast Fourier Transform (FFT) to I(t) and Q(t),two sets of frequency-domain signals,IFFT and QFFT,are obtained.
The complex expression of S21 parameter for KIDs under working bandwidth can be expressed as
After we get the PSD of KIDs,we only need to get the Responsivity of the resonance depth (A) of KIDs versus the input signal power (P),
2 Hardware design of homodyne mixing system
2.1 homodyne mixing system
The schematic diagram of the designed homodyne circuit is shown in
Figure 4.Schematic diagram of single channel homodyne mixing system circuit
The signal input segment of the system comprises a 15 dB directional coupler and a microwave switch A,alongside a 90 dB adjustable attenuation attenuator,tailored to accommodate the local oscillator port's 15dBm operating power. This setup ensures compatibility with the KIDs detectors,which typically handle signals below -80dBm. In the homodyne mixing architecture,symmetry in signal frequency is essential for IQ-mixer operation,with one signal entering the local oscillator port and the other,after power adjustment,entering the Dewar. However,IQ-mixer imperfections like amplitude and phase imbalances necessitate calibration,termed IQ-Calibration. This process demands signals at similar frequencies to enter the mixer's LO and RF ports,facilitated by a microwave switch toggling between KIDs measurement and calibration.
In signal processing circuit,it revolves around two LNA stages,a 0-60 dB adjustable attenuator,and a microwave switch,with the IQ-mixer as its linchpin. A 90 dB adjustable attenuator precedes the RF signal's entry into the Dewar refrigeration system,housing the KIDs samples. The signal through internal attenuators,KIDs,and the low-temperature amplifier HEMT before undergoing amplification through the LNAs and attenuators. A 60 dB adjustable attenuator interposed between LNA stages serves two critical functions: preventing signal saturation in subsequent LNAs and ensuring linear operation of the IQ mixer by controlling input power. The signal generator feeds the LO port,while KIDs' signal passes through LNA and attenuators before mixing,ideally generating two signals with matched amplitudes and a 90-degree phase difference.
Finally,the signal output and acquisition circuitry entail low-pass filters and impedance matching,dividing the IQ mixer's output into I and Q channels. These signals,along with ADC port connections,are managed through SMA lines and impedance matching. The ADC captures I(t) and Q(t) signals in the time domain for further processing.
Microwave switches A and B are devices of the same model,which can be controlled by an electronic control switch to switch between three working modes.<1>. 'Homodyne mixing mode',when A switch is connected to 1 and B switch is connected to 1,homodyne forms a complete loop to measure the characteristics of KIDs;<2>.In the 'IQ-mixer calibration mode',when the A switch is connected to 2 and the B switch is connected to 1,both the input and output Dewar ports are directly connected using SMA wires to form a calibration circuit. <3>. In the 'VNA mode',when the A switch is connected to 2 and the B switch is connected to 2,the signal no longer passes through the mixer,but only the two-stage amplifier LNA and attenuator can be connected to VNA to measure the S21 parameters of KIDs.
2.2 KIDs Noise Measurement Based on VNA
when measuring the noise of KIDs to obtain their PSD,it is usually achieved using the homodyne mixer system. After studying the existing new VNA,we found that it can also achieve noise measurement for KIDs. We use Keysight's model P5003A VNA,which can achieve vector network division mode,signal source mode,spectrometer mode,and CW mode (time domain sampling).
The hardware connection settings are as shown in section 2.1 of the 'VNA mode'. VNA software is set to CW mode,setting the frequency to be tested,sampling duration,and observation points. The specific settings are shown in
Figure 5.Setup window for VNA noise measurement in CW mode
Figure 6.Time domain interface between I and Q channels of VNA
It should be noted that the observation points set by the VNA UI program are not the actual sampling points. The actual sampling points are related to the settings of VNA's' IF Bandwidth ',' Dwelltime ',sampling duration,and observation points (set by the instrument development company program). The sampling rate and sampling points of VNA cannot be set by the user,but can be calculated using formulas (
NSET | T(ms) | TDwell-time(us) | Actual sampling frequency(kHz) |
---|---|---|---|
20 000 | 100 | 1 | 250 |
20 000 | 300 | 11 | 250 |
20 000 | 1 000 | 46 | 250 |
40 000 | 200 | 1 | 250 |
Table 1. Sampling frequency for different VNA settings
Design of dual channel homodyne based on cross correlation operation
When we use a single channel homodyne mixing system based on Chapter 2.1 to characterize their characteristics,there are the following problems: 1. The noise collected by existing measurement systems is not only the noise of KIDs,but also the superposition of noise from amplifiers,attenuators,directional couplers,mixers,and other components of the KIDs readout system. KIDs is ultra-high sensitivity detectors,so noise generated by KIDs is lower than noise generated by some devices in the readout system will have a certain impact on measurement accuracy. 2. The measurement system will introduce external interference signals(such as 50 Hz frequency doubling signals,ADC interference signals,etc.) generated by other than KIDs. The common noise measurement systems cannot effectively solve this problem.
To solve the above problems,we used two single channel homodyne mixing systems combined with hardware testing methods of power splitters,as well as software cross correlation algorithms,to effectively suppress the noise generated by the homodyne mixing system and to some extent suppress interference signals. The specific hardware circuit is shown in
Figure 7.Schematic diagram of dual channel homodyne mixing system circuit
The signal generator outputs a signal,which is divided into two channels by the directional coupler and these two signals enter into LO port of homodyne mixing system (LO1 and LO2). HMET amplifies the output signal of KIDs in the Dewar,and divides the output signal of HEMT into two identical signals as RF1 and RF2 through a power divider outside the Dewar. RF1 and LO1,as well as RF2 and LO2,are used as input signals for No.1 and No.2 homodyne mixing system,respectively. After output signals of two homodyne mixing system is mixed by IQ-mixer,the two time-domain signals are inputted into the ADC through an impedance converter. The signals are sampled as four ADC acquisition channels of dual channels I1 (t) and Q1 (t),I2(t) and Q2(t). Finally,the two-time domain signals are converted into amplitude or phase noise in the frequency domain through Autocorrelation and Fourier transform.
3 Software Algorithm Processing
3.1 IQ mixer calibration
The two signals generated by an ideal IQ-mixer with the same amplitude and a phase difference of 90 ° can be represented by functions such as formulas (
Actually,the differences in device properties of IQ-mixer with RF frequency cause 3 problems: 1. The amplitude imbalance; 2. The phases imbalance 3. DC offset (between I and Q channels). This will cause errors in the noise measurement of KIDs. To eliminate this error,it is necessary to first correct the imbalance of the IQ-mixer,obtain the correction parameters for the imbalance,and then use these parameters to calibrate the KIDs measurement data at the corresponding frequency points to obtain accurate IQ two-way microwave signals.
According to section 2.1,set the homodyne mixing system to the ' IQ-mixer calibration mode '. Two signal generators (one signal provided by the VNA of signal source mode) generate microwave signals close to the same frequency,one of which serves as an RF signal and the other as a local oscillator signal. The frequency settings of the two signal generators differ by a few kHz. Then use outputs data to draw an output circle (actually an ellipse before calibration),which is represented by the following two equations:
Among them,I0 and Q0 (the center of the ellipse) represent the offset of the I and Q channel outputs,respectively; AI/AQ (the ratio of the major axis of an ellipse to its minor axis) which is represented amplitude difference; γ which is the inclination angle of the ellipse,represents a fixed initial phase difference. The least squares ellipse fitting is introduced here to calibrate the imbalance of the IQ-mixer. The fitting method is given by the following equation(
The five main parameters of the fitted ellipse,I0,Q0,AI,AQ,γ It can be represented by fitting parameters as formula(
To obtain an ellipse fitted due to imbalance,we first need to correct the five parameters obtained in the previous section of the ellipse to a standard circle. Firstly,we deduct the DC offset (I-I0,Q-Q0) and return the corresponding ellipse center to the coordinate origin. Then calibrate the phase imbalance using the coordinate transformation formula shown in
Figure 8.Diagram of coordinate rotation formula
Figure 9.Schematic diagram of IQ mixer calibration complex frequency domain
3.2 7 parameter fitting
Due to the impedance mismatch of the test link,standing wave,signal transmission delay of the cable,and the quality factor of KIDs themselves,the resonant characteristics of KIDs collected by ADC are different from the ideal resonant characteristics. According to the fitting data,the influence of these parameters on the resonant curve of KIDs can be eliminated. The equation of the S21 curve of the general KIDs resonator is
Where f is the detection frequency,fr is the resonant frequency,Ql is the load quality factor,and Qc is the absolute value of the coupling quality factor,φ Indicates impedance mismatch,a is additional amplitude,α is phase shift,And electronic delay caused by cable length and limited speed of light τ。 The data collected by ADC was fitted with seven parameters using the above formula,and then calibrated using
3.3 Cross correlation algorithm
According to the content in Section 2.3,the two-channel homodyne mixing system is used to sample the time domain data of the two channel I and Q channels through ADC,and four channel data of I1 (t),Q1 (t),I2 (t) and Q2 (t) as shown in
Figure 10.Schematic diagram of cross correlation algorithm for dual channel homodyne mixing system
The cross-correlation operation Xcor can effectively represent the noise generated by the same KIDs measured by a dual channel homodyne mixing system,while the noise generated by the measurement system is suppressed by the cross-correlation operation,thereby enhancing the accuracy of KIDs noise (PSD) characterization in the measurement system.
4 Data processing and analysis
The Homodyne mixing system and KIDs settings are shown in the
Figure 11.The Homodyne mixing system and KIDs setup
4.1 Impact of IQ-calibration on IQ-Mixer linear interval
Figure 12.IQ Mixer linear area @ 2 GHz
Figure 13.IQ linear area @ 4 GHz
Figure 14.IQ Mixer linear area @ 6 GHz
The linear operating range of the IQ mixer is an important parameter for the entire homodyne system. When the KIDs output signals enter the linear range of the IQ Mixer,it is necessary to ensure that the signal does not experience distortion. We measured the input-output power curve of the IQ mixer by combining the signal source and spectrometer. We selected four frequency points of 2 GHz,4 GHz,6 GHz,and 8 GHz to test the corresponding RF input power/IQ output amplitude curves,where the output amplitude=20 log √ (I^2+Q^2). The measurement results are shown in Fig.
Figure 15.IQ Mixer linear area @ 8 GHz
After correction,the linear point of the IQ Mixer corresponding to the 2 GHz,4 GHz,6 GHz,and 8 GHz frequency points are 47 dB,46 dB,68 dB,and 56 dB,respectively. The uncalibrated curve has a narrow linear interval and singular inflection points. After calibration,the linear intervals are all greater than 45dB,with a significant increase in linear intervals and no singular inflection points. The saturation power of the system on the right side of the linear region is approximately 5dBm,and the output signal does not change as the input power increases. The bottom noise of the system is reached on the left side of the linear region,and does not change as the input signal strength decreases.
When the input power of the signal is too low,the output signal power will approach the DC offset. If the DC offset is negative and the curve monotonically increases or the DC offset is positive and the curve monotonically decreases,then the curve value will be squared and a turning point will appear. This explains the reason for the singular turning point in the linear interval of IQ Mixer corresponding to 2 GHz and 8 GHz frequency points. After IQ calibration,the DC offset is deducted and the inflection point does not exist.
4.2 Measurement of S21 parameters and seven parameter fitting
The S21 of KIDs measured by the homodyne mixing system is shown in
Figure 16.S21 of KIDs measured by homodyne mixing system
4.3 PSD
We conducted time series sampling on the resonant and non-resonant frequency points of KIDs resonators. The non resonance were used to characterize the overall noise level of the detection link,while the noise at the resonance was used to characterize the noise level of the detector. The collected noise is obtained from an ADC with a sampling rate of 200 kHz and 20000 sampling points.
We used the above method to perform temporal sampling on the resonance and non-resonance of KIDs,followed by IQ calibration and seven parameter fitting,and finally obtained the PSD of the detector through Fourier transform. Firstly,the time-domain signal of KIDs chips with a resonant point of 4 186 MHz was measured using a single channel homodyne mixing system. Then,IQ-calibration and seven parameter fitting were performed on the noise points of the resonant and non-resonance. It can be seen that the resonance noise collected in the time domain of the resonance is plotted on the far-left side of the resonance circle graph,while the non-resonance is plotted in the area of the baseline (x-axis approaching 1) on the right side of the graph,as shown in
Figure 17.Complex frequency domain diagram of KIDs resonance and non-resonance noise
Figure 18.PSD of KIDs on resonance and non-resonance
From the
Figure 19.Comparison of resonance PSD between single channel and dual channel homodyne system testing KIDs
Based on the content of Chapter 2.2,we measured the noise of the aforementioned KIDs chips using VNA and obtained their PSD diagrams,as shown in
Figure 20.KIDs PSD Test Result Graph Based on VNA
4.4 Responsivity and NEP
We obtained a curve of resonance depth with temperature by changing the temperature of the cryostat where KIDs are located,and fitted the values of dA/dT (3.5 * 10-7) as shown in
Figure 21.S21 parameters of KIDs at different temperatures and DA/dT fitting curve
Figure 22.NEP result of KIDs
5 Conclusion
This article designs a novel Homodyne mixing system to characterize the performance of KIDs. The newly designed single channel homodyne mixing hardware system facilitates IQ mixer calibration and integration with VNA through the combination of microwave switches and other microwave devices; By combining the CW mode of the new VNA,the measurement of KIDs noise by VNA has been achieved; By designing a dual channel homodyne mixing system and combining it with dual channel algorithms,the influence of interference signals has been effectively suppressed. The software algorithm has achieved the calibration of IQ-mixer imbalance,expanding the linear range of the mixer to no less than 45dB. These research results have been applied to characterize the performance of KIDs and have played an important role in the design of KIDs arrays.